50
A-96.250.351 / 230320
AMU Rescon
Program List and Explanations
4.3 Logger
The instrument is equipped with an internal logger. The logger data
can be downloaded to a PC using the built-in RS232 interface.
The logger can save approx. 1500 data records. The Records con
-
sists of: Date, time, alarms, measuring value, raw value (MΩ), case
temperature, flow.
4.3.1
Log Interval:
Select a convenient log interval. Consult the table be
-
low to estimate the max logging time. When the logging buffer is
full, the oldest data record is erased to make room for the newest
one (circular buffer).
Range: 1 Second to 1 hour
4.3.2
Clear Logger:
If confirmed with
yes
, the complete logger data is de
-
leted. A new data series is started.
5 Installation
5.1 Sensors
5.1.1
Flow:
Select the type of flow cell. Possible flow cells:
Selection of the suitable flow cell.
5.1.2
Meas. Mode:
The two measuring modes Conductivity or Resistivity
are available.
Interval
1 s
5 s
1 min
5 min
10 min
30 min
1 h
Time
25 min
2 h
25 h
5 d
10 d
31 d
62 d
Mode
B-Flow
Q-Flow
QV-flow
None
Q-flow
Q-Hflow
Flow cell type
Mode
B-Flow
None
Q-Flow or QV-Flow
Q-flow
Q-Hflow or QV-Hflow
Q-Hflow