46
A-96.250.331 / 230320
AMU Powercon
Program List and Explanations
The value is simulated by the relay/signal output.
At the absence of any key activities, the instrument will switch back
to normal mode after 20 min. If you quit the menu, all simulated val
-
ues will be reset.
3.3 Set Time
Adjust date and time.
3.3 Quality Assurance
3.4.5
Follow the commands on the screen. Save the value with the
<Enter> key.
4 Operation
4.1 Sensors
4.1.1
Filter Time Constant:
Used to damp noisy signals. The higher the
filter time constant, the slower the system reacts to changes of the
measured value.
Range: 5–300 Sec
4.1.2
Hold after Cal.:
Delay permitting the instrument to stabilize again af
-
ter calibration. During calibration plus hold-time, the signal outputs
are frozen (held on last valid value), alarm values, limits are not ac
-
tive.
Range: 0–6‘000 Sec
4.2 Relay Contacts
4.3 Logger
The instrument is equipped with an internal logger. The logger data
can be downloaded to a PC using the built-in RS232 interface.
The logger can save approx. 1500 data records. The Records con
-
sists of: Date, time, alarms, measured value, measured value un
-
compensated, temperature, flow.
Range: 1 second to 1 hour
Alarm Relay:
Active or inactive.
Relay 1 and 2:
Active or inactive.
Signal Output 1 and 2:
Actual current in mA