Program List and Explanations
8.
Program List and Explanations
1 Messages
1.1 Pending Errors
1.1.5
Provides the list of active errors with their status (active, acknowl
-
edged). If an active error is acknowledged, the alarm relay is active
again. Cleared errors are moved to the Message list.
1.2 Message List
1.2.1
Shows the error history: Error code, date / time of issue and status
(active, acknowledged, cleared). 65 errors are memorized. Then
the oldest error is cleared to save the newest error (circular buffer).
2 Diagnostics
In diagnostics mode, the values can only be viewed, not modified.
2.1 Identification
Desig.
:
Designation of the instrument.
Version
: Firmware of the instrument (e.g. V6.20-07/16)
2.1.4
Factory Test
: Test date of the instrument, motherboard and
frontend
2.1.5
Operating Time:
years / days / hours / minutes / seconds
2.2 Sensors
2.2.1
Cond. Sensor
Current value
in µS
Raw value
in µS
Cell Constant
2.2.1.4
QA History: Review the QA values (Number, Date-Time, Deviation
Conductivity, Deviation Temperature) of the last quality assurance
procedures. Only for diagnostic purpose. Max. 65 data records are
memorized.
2.2.1.5
Cal. History
: Review diagnostic values of the last calibrations.
Only
for diagnostic purpose.
Number; Date, Time
Cell constant
Max. 64 data records are memorized. One process calibration cor
-
responds to one data record.