Default Values
11.
Default Values
Operation:
Sensors:
Filter Time Const.:..................................................................10 Sec
Hold after Cal.:..................................................................... 300 sec
Relay Contacts
Alarm Relay .................................................. same as in Installation
Relay 1/2 ...................................................... same as in Installation
Input.............................................................. same as in Installation
Logger:
Logger Interval:.............................................................. 30 Minutes
Clear Logger: ................................................................................no
Installation:
Sensors
Flow: ........................................................................................ None
Sensor Parameters; Cell Constant ...............................0.0415 cm
-1
Sensor Parameters; Temp. corr. ...........................................0.00 °C
Sensor Parameters; Cable length............................................0.0 m
Sensor Parameters; Meas. unit ............................................. µS/cm
Temp. Compensation; Comp. ...................................................none
Quality Assurance; Level ......................................................... 0: Off
Signal Output 1
Parameter: .................................................................... Conductivity
Current loop: .....................................................................4 –20 mA
Function: .................................................................................. linear
Scaling: Range low: ........................................................... 0.000 µS
Scaling: Range high:................................................................ 1 mS
Signal Output 2
Parameter: ....................................................................Temperature
Current loop: .....................................................................4 –20 mA
Function: .................................................................................. linear
Scaling: Range low: ...................................................................0 °C
Scaling: Range high:................................................................50 °C
Alarm Relay:
Alarm Conductivity:
Alarm high:........................................................................... 300 mS
Alarm low: .......................................................................... 0.000 µS
Hysteresis: ........................................................................... 1.00 µS
Delay:.......................................................................................5 Sec
Sample Flow:
Flow Alarm.................................................................................. yes
Alarm high:.............................................................................. 20 l/h
Alarm low: ................................................................................. 5 l/h
Sample Temp:
Alarm High: ............................................................................160 °C
Alarm Low:.................................................................................0 °C