![Swann AMI Deltacon Power Скачать руководство пользователя страница 46](http://html1.mh-extra.com/html/swann/ami-deltacon-power/ami-deltacon-power_operators-manual_1411825046.webp)
A-96.250.451 / 190121
AMI Deltacon Power
Maintenance
46
6.8. Longer Stop of Operation
1
Stop sample flow.
2
Slightly squeeze the ion exchanger bottle.
Thus no water will spill out of the flow cell when loosening the
bottle.
3
Unscrew and carefully remove the ion exchanger bottle with the
exhausted resin.
4
Close the ion exchanger bottle with the screw cover and store it
in a frost-protected room.
5
Screw an empty bottle into the bottle holder.
6
Shut off power of the instrument.