1.3. Restriction for use
The AMI Deltacon DG is designed for determination of:
specific (total) conductivity
cation (acid) conductivity after the cation exchanger
degassed conductivity after a sample reboiler in water-steam-
cycles only.
It calculates the pH value and the concentration of the alkaline sub
-
stance (NH
3
, Morpholine, etc.) if an alkaline substance is present in
the water.
It is not suitable for pH determination in high purity water before
alkalization reagent addition.
Conditions for pH calculation:
only 1 alkalization agent in the sample
the contamination is mostly NaCl
phosphate concentration is < 0.5 ppm
pH value is > 7.5, and < 11.5
No sand. No oil.
The sample must not contain any particles, which may block the flow
cell. Sufficient sample flow is coercive for the correct function of the
instrument.