AMI Deltacon DG
Default Values
80
11. Default Values
Operation:
Sensors:
Filter Time Const.: ........................................................................20 s
Hold after Cal.:................................................................................0 s
Relay Contacts
Alarm Relay.....................................................same as in Installation
Relay 1/2.........................................................same as in Installation
Input.................................................................same as in Installation
Logger:
Logger Interval:........................................................................ 30 min
Clear Logger:.................................................................................. no
Display:
Screen 1 and 2; Row 1:.................................................... Cond 1(sc)
Screen 1 and 2; Row 2:.....................................................Cond 2(cc)
Screen 1 and 2; Row 3:..............................................................None
Installation:
Sensors
Miscellaneous; Calculations: .......................................................... no
Miscellaneous; Meas. unit ........................................................µS/cm
Miscellaneous; Monitoring of resin ................................................. no
Miscellaneous; Resin Capacity:..................................................... 1.8
Miscellaneous; Volume of resin:...................................................1.0 l
Sensor Parameters; Sensor 1, 2 and 3;Cell Constant.... 0.0415 cm
-1
Sensor Parameters; Sensor 1, 2 and 3; Temp. corr. .............. 0.00 °C
Sensor Parameters; Sensor 1, 2 and 3; Cable length............... 0.0 m
Sensor Parameters; Sensor 1; Temp. comp.; Comp: .......... Ammonia
Sensor Parameters; Sensor 2; Temp. comp.; Comp: ..... Strong Acids
Sensor Parameters; Sensor 3; Temp. comp.; Comp: ..... Strong Acids
Flow:..............................................................................................yes
Degasser; Mode ............................................................................. on
Degasser; Boiling point interval..................................................... 6 h
Signal Output 1
Parameter:.........................................................................Cond 1(sc)
Current loop:....................................................................... 4 –20 mA
Function:.....................................................................................linear
Scaling: Range low:..............................................................0.000 µS
Scaling: Range high: ........................................................1000.00 µS
Signal Output 2
Parameter:.........................................................................Cond 2(cc)
Current loop:....................................................................... 4 –20 mA
Function:.....................................................................................linear
Scaling: Range low:..............................................................0.000 µS
Scaling: Range high: ........................................................1000.00 µS