
A-96.250.491 / 190121
AMI Deltacon DG
Maintenance
42
6.6. Longer Stop of Operation
1
Stop sample flow.
2
Slightly squeeze the ion exchanger bottle.
Thus no water will spill out of the flow cell when loosening the
bottle.
3
Unscrew and carefully remove the ion exchanger bottle with the
exhausted resin.
4
Close the ion exchanger bottle with the screw cover and store it
in a frost-protected room.
5
Screw an empty bottle into the bottle holder.
6
Shut off power of the instrument.