A-96.250.491 / 190121
AMI Deltacon DG
Instrument Setup
31
4.
Instrument Setup
After the analyzer is installed according to the previous instructions,
connect the power cord. Do not switch on power, yet!
4.1. Establish sample flow
1
Open flow regulating valve, see
.
2
Wait until the flow cell has been completely filled.
3
Switch on power.
4
Adjust the sample flow to 5 - 10 l/h.
5
Let the instrument run-in for 1 h.
This recommendation is valid for rinsed cation exchanger
resin (nuclear grade) delivered by Swan.
Note:
Not rinsed cation exchanger resin from other suppliers can
take a run-in period of several hours to several days.
4.2. Programming
Sensor
parameters
Program all sensor parameters in Menu Installation-Sensors:
menu 5.1.2.1, sensor 1
menu 5.1.2.2, sensor 2
menu 5.1.2.3, sensor 3
The sensor characteristics are printed on the label of each sensor.
Enter for each sensor separately the:
Cell constant [cm
-1
]
Temperature correction [°C]
Note:
Cable length [m] Set the cable length to 0.0 m if the
sensors are installed in the flow cell of the AMI monitor.
Temperature compensation: The default setting for sensor 1
(specific conductivity) is ammonia.
87-344.203
UP-Con1000SL
Sensor type
SW-xx-xx-xx ZK = 0.0417
Cell constant
SWAN AG
DT = 0.06 °C
Temperature correction