AMI Deltacon DG
Installation
24
3.5.2
Power Supply
The mains supply is connected to the degasser controller. From
there, the AMI transmitter and the heater are supplied.
WARNING
Electrical shock hazard
Installation and maintenance of electrical parts must be per
-
formed by professionals. Always turn off power before manipulat
-
ing electric parts.
Mains
connection
Note:
The protective earth wire (ground) has to be connected to
the grounding terminal.
Installation
requirements
The installation must meet the following requirements.
Mains cable complies with standards IEC 60227 or IEC 60245;
flammable rating FV1
Mains equipped with an external switch or circuit-breaker
– near the instrument
– easily accessible to the operator
– marked as interrupter for AMI Deltacon DG
D
E
Already connected
external devices:
Transmitter supply
Heater supply
A
B
E
C
D
A
B
C
Neutral conductor
Phase conductor
Protective earth