AMI Codes-II
Program List and Explanations
3 Maintenance
3.1 Calibration
In this menu, you can correct measuring values (all disinfectants
and pH) or calibrate offset and slope of pH electrode.
3.1.1
Process DIS
: Possibility to correct the disinfectant value.
See
Process Calibration of DIS, p. 50
, for more details.
3.1.2
Process pH
: Only available, if option pH has been installed.
Correction of pH electrode. See
Process Calibration of DIS, p. 50
for details.
3.1.3
Standard pH:
Only available, if option pH has been installed.
Calibration of pH electrode with the two standard solutions
programmed in Installation 5.1.3. See
, for
details.
3.2 Service
3.2.1
Verification: Performs a verification using the reference kit. Follow
dialog. See
3.2.2
Fill System: Activates the reagent pump.
3.3 Simulation
To simulate a value or a relay state, select the
alarm relay,
relay 1or 2
signal output 1 or 2
with the [
] or [
] key.
Press the [Enter] key.
Change the value or state of the selected item with the [
] or
[
] key.
Press the [Enter] key.
The value is simulated by the relay/signal output.
At the absence of any key activities, the instrument will switch back
to normal mode after 20 min. If you quit the menu, all simulated val-
ues will be reset.
Alarm Relay:
Active or inactive
Relay 1 and 2:
Active or inactive
Signal Output 1 and 2:
Actual current in mA
Signal Output 3:
Actual current in mA (if option is installed)