![Swann AMI Codes-II Скачать руководство пользователя страница 72](http://html1.mh-extra.com/html/swann/ami-codes-ii/ami-codes-ii_operators-manual_1411948072.webp)
70
A-96.250.571 / 050517
AMI Codes-II
Program Overview
8.
Program Overview
For explanations about each parameter of the menus see
.
Menu 1
Messages
informs about pending errors and mainte-
nance tasks and shows the error history. Password protection
possible. No settings can be modified.
Menu
2
Diagnostics
is always accessible for everybody. No
password protection. No settings can be modified.
Menu 3
Maintenance
is for service: Calibration, simulation of
outputs and set time/date. Please protect with password.
Menu 4
Operation
is for the user, allowing to set limits, alarm
values, etc. The presetting is done in the menu Installation
(only for the System engineer). Please protect with password.
Menu 5
Installation
: Defining assignment of all inputs and
outputs, measuring parameters, interface, passwords, etc.
Menu for the system engineer. Password strongly recom-
mended.
8.1.
Messages (Main Menu 1)
8.2.
Diagnostics (Main Menu 2)
Pending Errors
Pending Errors
1.1.5*
* Menu numbers
1.1*
Maintenance List
Maintenance List
1.2.5*
1.2*
Message List
Number
1.3.1*
1.3*
Date, Time
Identification
Designation
AMI Codes-II
* Menu numbers
2.1*
Version
V6.20 - 08/16
Peripherals
PeriClip 1 / 1.03
2.1.3.1*
2.1.3*
PeriClip 2
only with cleaning module
Factory Test
Instrument
2.1.4.1*
2.1.4*
Motherboard