AMI Codes-II CC
Program List and Explanations
5.3.1.1.2
Total chlorine 1
5.3.1.1.2.1
Alarm High:
If the measured value rises above the alarm high val-
ue, the alarm relay is activated and E003 is displayed in the mes-
sage list. Range: 0.00–10.00 ppm
5.3.1.1.2.26
Alarm Low:
If the measured value falls below the alarm low value,
the alarm relay is activated and E004 is displayed in the message
list.
Range: 0.00–10.00 ppm
5.3.1.1.2.36
Hysteresis:
Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value fluc-
tuates around the alarm value.
Range. 0.00–10.00 ppm
5.3.1.1.2.46
Delay:
Duration, the activation of the alarm relay is retarded after
the measuring value has risen above/fallen below the programmed
alarm.
Range: 0.00–28‘800 Sec
5.3.1.1.3
Total chlorine 2
5.3.1.1.3.1
Alarm High:
If the measured value rises above the alarm high val-
ue, the alarm relay is activated and E005 is displayed in the mes-
sage list. Range: 0.00–10.00 ppm
5.3.1.1.3.26
Alarm Low:
If the measured value falls below the alarm low value,
the alarm relay is activated and E006 is displayed in the message
list.
Range: 0.00–10.00 ppm
5.3.1.1.3.36
Hysteresis:
Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value fluc-
tuates around the alarm value.
Range. 0.00–10.00 ppm
5.3.1.1.3.46
Delay:
Duration, the activation of the alarm relay is retarded after
the measuring value has risen above or fallen below the pro-
grammed alarm.
Range: 0.00–28‘800 Sec