AMI Codes-II CC
Maintenance
A-96.250.581 / 220616
57
6.8.1
Disassemble the Flow Cell
The flow cell can be disassembled easily. Before disassemble the
flow cell, switch off the instrument according to instructions in
of Operation for Maintenance, p. 45
.
Cleaning
1
Switch off the instrument according to instructions in
Operation for Maintenance, p. 45
2
Remove the constant head cover [A].
3
Remove the constant head tube [C] from the flow cell block.
4
Pull the overflow tube [B] out of the flow cell block [D].
5
Clean all acrylic parts with a soft brush (bottle cleaner) and
soapy water.
6
Remove lime deposits with a common household deliming
agent with standard concentrations.
7
Rinse the parts carefully with clean water.
A
B
C
D
E
Constant head cover
Overflow tube
Constant head tube
Flow cell block
Flow regulating valve
Level
A
B
D
E
C