A-96.250.511 / 091120
AMI Turbiwell
Troubleshooting
79
7.5. Replacing Fuses
WARNING
External Voltage.
Externally supplied devices connected to relay 1 or 2 or to the
alarm relay can cause electrical shocks
Make sure that the devices connected to the following contacts
are disconnected from the power before resuming installation.
– relay 1
– relay 2
– alarm relay
Find and repair the cause for the short circuit before replacing the
fuse.
Use tweezers or needle-nosed pliers to remove the defective fuse.
Use original fuses provided by SWAN only.
A
B
C
D
E
F
G
AC variant: 1.6 AT/250 V Instrument power supply
DC variant: 3.15 AT/250 V Instrument power supply
1.0 AT/250V Relay 1
1.0 AT/250V Relay 2
1.0 AT/250V Alarm relay
1.0 AF/125V Signal output 2
1.0 AF/125V Signal output 1
1.0 AF/125V Signal output 3