A-96.250.671 / 070222
AMI Silica
Program List and Explanations
81
4.4 Logger
The instrument is equipped with an internal logger. The logger data
can be copied to a PC with an USB stick if option USB interface is in
-
stalled.
The logger can save approx. 1500 data records. Records consists of:
Date, time, alarms, measured value, temperature, flow.
Range: 1 second to 1 hour
4.4.1
Log Interval:
Select a convenient log interval. Consult the table be
-
low to estimate the max logging time. When the logging buffer is full,
the oldest data record is erased to make room for the newest one
(circular buffer).
Event driven or interval (see table below)
4.4.2
Clear Logger:
If confirmed with <yes>, the complete logger data is
deleted. A new data series is started.
5 Installation
5.1 Sensors
5.1.1
Ref. Verification:
Set absorbance value of verification kit according to
label.
Range: 0.150–0.600
5.1.2
Standard:
Set the value in ppb of the standard you use for calibra
-
tion.
Range: 50 ppb–5 ppm
5.1.3
Blank
: If the reagents contain a known amount of Silica, “Blank” can
be used to shift the zero point. The entered Blank value then is sub
-
tracted from the measuring value.
Range: 0.0 ppb–10.0 ppb
5.1.4
Meas. Interval
: The measuring interval can be set to 10, 15, 20 or 30
minutes.
5.1.5
Channels:
If the 2
nd
sample stream option is installed, you can choose 1 or 2
channels.
If an AMI Sample Sequencer is connected to the AMI Silica, you can
choose 1 to 6 channels.
Interval
1 s
5 s
1 min
5 min
10
min
30
min
1 h
Event
Driven
Time
25 min 2 h
25 h
5 d
10 d
31 d
62 d