A-96.250.671 / 070222
AMI Silica
Maintenance
50
Reagent
consumption
The 2 liter reagent canister will last for 1 month of operation with de
-
fault measurement interval of 10 minutes. The provided reagent set
for 3 canister fillings therefore lasts for 3 months of operation.
Note:
Please consider the following two points when preparing
new reagents:
• Reagent 3, Oxalic Acid dissolves very slowly, we therefore
recommend to prepare Reagent 3 first.
• Reagent 1, add Sodium hydroxide (Reagent 1b) first.
General
procedure
1
Rinse all containers well with demineralized water.
2
Fill the container to 3/4 of its final level with demineralized water.
3
Carefully add the chemicals. See
,
,
4
Screw the cover onto the container and mix well.
5
Fill the container to its final level and mix again.
6
Put container into the holder, number 1 to 4 from left to right.
7
Insert suction lances into the containers; make sure that the num
-
bers on the suction lances correspond to the numbers on the
containers.
8
Lock cover.
Measuring interval
Duration per
canister
Duration per
reagent set
10 minute
~ 1 month
3 months
15 minutes
~ 1.5 months
4.5 months
20 minutes
~ 2 months
6 months
30 minutes
~ 3 months
9 months