6.3. Refill or replace Reagents
The liquid level in container 4 is monitored. The following messages
are displayed.
WARNING
Health hazard
For safe handling of the reagents you must read and under
-
stand the Material Safety Data Sheets (MSDS).
Only persons trained in handling dangerous chemicals are al
-
lowed to prepare the reagents.
Canister setup
Container almost
empty
Maintenance E065 - Reagents low and the
remaining reagent volume in % (starting at 17%
= 340 ml).
Container empty Error E022 - Reagent empty
A
B
C
D
E
F
G
H
I
Suction lance without
level detector
(containers 1–3)
Suction lance with
level detector
(container 4)
Level detector
2 L mark
Container 1:
Ammonium molybdate
Container 2:
Sulfuric Acid
Container 3:
Oxalic Acid
Container 4:
Ammonium Ferrous
Sulfate
Holder