46
A-96.250.631 / 280219
AMI Codes-II TC
Maintenance
6.3.
Refill or replace Reagents
The liquid level in canister 2 is monitored. The following messages
are displayed:
NOTICE:
Before refilling the reagents, rinse the canisters with
demineralized water.
Reagent
consumption
The reagent consumption depends on the selected measuring in-
terval and on the selected operating mode (measuring range: Stan-
dard / Extended). Below you will find some examples:
The provided reagent set is sufficient for 8 canister fillings.
canister almost
empty
Maintenance E065 - Reagents low and the
remaining reagent volume in %
(starting at 17 % = 340 ml).
canister empty
Error E022 - Reagent empty
Measuring
interval
Duration per canister
(measuring range:
standard)
Duration per canister
(measuring range:
extended)
3 minutes
13–19 days
8–11 days
5 minutes
22–32 days
13–19 days
10 minutes
44–63 days
26–38 days
20 minutes
88–126 days
53–76 days