36
SVAN 979 User Manual
Note: The calibration factor is always added to measurement results and measurement range limits
of the Level Meter, 1/1 Octave, 1/3 Octave and other functions.
Note: The calibration level and the calibration result are expressed in different units depending on
the settings of the instrument. The metric or non-metric Vibration units are set in the Vibration Units
screen (path: <Menu> / Auxiliary Setup / Vibration Units). Additionally, the linear or logarithmic units
are set in the Display Scale screen (path: <Menu> / Display / Display Scale).
Note: It is not possible to check and calibrate the instrument during the execution of live
measurements. It is possible to open different lists and sub-lists but the positions in these lists are
displayed greyed out inversely and so -
not accessible. The flashing „measurement” icon on the top
line indicates that the instrument is in the measurement process. To check or calibrate the
instrument, the current measurement in progress must be finished!
3.3.1
System Check
ISO 8041-1:2017 standard advises users to perform in-situ checks of measuring
instruments. Checking should be carried out immediately before and after the
measurement.
1. Select
System Check
in the
Calibration
sub-list and press
<ENTER>
.
2. Set the reference calibration level (
Cal. Level
)
– see Appendix C,
Chapter C.1, par. Calibration.
3.
Attach the sound calibrator to the instrument’s microphone or accelerometer
to the vibration calibrator.
4. Switch on the calibrator and wait approximately 30 seconds before starting
the system check measurement.
5. Start the calibration measurement with the
<Start>
key.
The measurement starts without delay. If maximal difference between three 1-
second consecutive results (Leq(C) or RMS) is less than
0.05dB
, the
measurement will be stopped. Otherwise the user should stop it manually.
During the system check measurement, the
<ESC>
and
<Pause>
keys do not
operate but it is possible to stop the measurement using the
<Start/Stop>
key.
System check is considered as successful if its result (
Cal. Result
) does not
differ from the calibration level by more than 0.5 dB.
If system check measurement shows bigger difference than 0.5 dB you should
perform
Calibration by Measurement
.
6. Press
<ENTER>
to exit
System Check
.
Note: It is advised to perform the system check of the instrument each time before the
measurements begin. If system check gives negative result, then it is necessary to perform
calibration.
3.3.2
Calibration by Sensitivity (Acoustic signal)
Calibration on the base of the microphone’s
published sensitivity information can be
performed the following way:
1. Select
the
Calibr. by Sensitivity
position in the
Calibration
list and press
<ENTER>
.
<ENT>
Содержание SVAN 979
Страница 272: ...272 SVAN 979 User Manual Appendixes 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 273: ...273 SVAN 979 User Manual Appendixes 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 274: ...274 SVAN 979 User Manual Appendixes 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 275: ...275 SVAN 979 User Manual Appendixes 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 302: ...302 SVAN 979 User Manual Appendixes 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter ...
Страница 303: ...303 SVAN 979 User Manual Appendixes 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter ...
Страница 304: ...304 SVAN 979 User Manual Appendixes 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter ...
Страница 305: ...305 SVAN 979 User Manual Appendixes 16 0 Hz 1 1 octave filter 8 00 Hz 1 1 octave filter 4 00 Hz 1 1 octave filter ...
Страница 306: ...306 SVAN 979 User Manual Appendixes 2 00 Hz 1 1 octave filter 1 00 Hz 1 1 octave filter Summary curve ...
Страница 312: ...312 SVAN 979 User Manual Appendixes Summary curve ...
Страница 348: ...348 SVAN 979 User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 349: ...349 SVAN 979 User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz ...
Страница 350: ...350 SVAN 979 User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz ...
Страница 351: ...351 SVAN 979 User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz ...
Страница 352: ...352 SVAN 979 User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz ...
Страница 370: ...370 SVAN 979 User Manual Appendixes O ring gasket locations inside ...