threshold level for PTC calculation in profile 3, is set to 115 dB (XC115:3);
•
threshold level for ULT calculation in profile 1, is set to 115 dB (Xl115:1);
•
threshold level for ULT calculation in profile 2, is set to 115 dB (Xl115:2);
•
threshold level for ULT calculation in profile 3, is set to 115 dB (Xl115:3);
•
logger splitting is disabled (XA0);
•
logger splitting time 1 is disabled (XD-1:1);
•
logger splitting time 2 is disabled (XD-1:2);
•
logger splitting time 3 is disabled (XD-1:3);
•
logger splitting time 4 is disabled (XD-1:4);
•
logger splitting time 5 is disabled (XD-1:5);
•
logger splitting time 6 is disabled (XD-1:6);
•
instrument is in the Stop state (S0);
•
logger is active (T1);
•
exposition time is set to 8 hours (e480);
•
criterion level in profile 1 is chosen as 80 dB (c1:1);
•
criterion level in profile 2 is chosen as 80 dB (c1:2);
•
criterion level in profile 3 is chosen as 80 dB (c1:3);
•
threshold level in profile 1 is None (h0:1);
•
threshold level in profile 2 is None (h0:2);
•
threshold level in profile 3 is None (h0:3);
•
exchange rate in profile 1 is set to 3 (x3:1).
•
exchange rate in profile 2 is set to 3 (x3:2).
•
exchange rate in profile 3 is set to 5 (x5:3).
•
measurement trigger mode is switched off (m0);
•
LEQ
result from the first profile is used as the measure triggering signal (s0);
•
measurement trigger level is set to 100 dB (l100);
•
gradient in measurement trigger is equal to 10 dB/ms (O10)
•
LEQ
result from the first profile for
1/1 Octave
is used as the measurement trigger signal (o0);
•
LEQ
result from the first profile for
1/3 Octave
is used as the measurement trigger signal (t0);
Note: All bytes of that transmission are ASCII characters.
A.3 F
UNCTION
#2
-
MEASUREMENT RESULTS READ
-
OUT IN THE
SLM
MODE
#2 function enables one to read out the current measurement result from the selected profile.
#2 function
has the format defined as follows:
#2 [,<aver>] [,<profile>] [[[ ,X? ] ,X? ] ,(...) ];
where:
<aver>
– type of results:
i
– instantaneous results, i.e. results from the current cycle (default),
a
– averaged results, i.e. results from the previous cycle.
<profile>
– profile number:
1, 2 or 3
– one of the profile, i.e. only results from the given profile will be sent;
X
– code of the specified result (see below); if no code are specified all results will be sent;
Содержание SVAN 971
Страница 98: ...SVAN 971 User Manual 98 ...
Страница 154: ...SVAN 971 User Manual Appendixes 154 1000 Hz 1250 Hz 400 Hz 500 Hz 630 Hz 800 Hz ...
Страница 155: ...SVAN 971 User Manual Appendixes 155 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 156: ...SVAN 971 User Manual Appendixes 156 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 157: ...SVAN 971 User Manual Appendixes 157 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 169: ...SVAN 971 User Manual Appendixes 169 2 0 kHz 1 1 octave filter 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter ...
Страница 170: ...SVAN 971 User Manual Appendixes 170 250 Hz 1 1 octave filter 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter ...
Страница 177: ...SVAN 971 User Manual Appendixes 177 B filter Class 1 according to the IEC 651 B filter characteristics ...
Страница 206: ...SVAN 971 User Manual Appendixes 206 400 Hz 500 Hz 630 Hz 800 Hz 1000 Hz 1250 Hz ...
Страница 207: ...SVAN 971 User Manual Appendixes 207 1600 Hz 2000 Hz 2240 Hz 2500 Hz 2800 Hz 3150 Hz ...
Страница 208: ...SVAN 971 User Manual Appendixes 208 3550 Hz 4000 Hz 4500 Hz 5000 Hz 5600 Hz 6300 Hz ...
Страница 209: ...SVAN 971 User Manual Appendixes 209 7100 Hz 8000 Hz 8500 Hz 9000 Hz 9500 Hz 10000 Hz ...
Страница 220: ...SVAN 971 User Manual Appendixes 220 ...