SVAN 973 User Manual
26
3 MEASUREMENT FUNCTIONS AND CALIBRATION
– Function
In the
Function
section, you can select the measurement
function (
Meas. Funct
) and perform the instrument
calibration (
Calibration
).
To open the
Function
section, press the
<Menu>
key,
select the
Function
position and press
<Enter>
.
<ENT>
3.1 A
CTIVATING MEASUREMENT FUNCTIONS
–
M
EASUREMENT
F
UNCTION
The main function of the instrument is measurements of the broadband sound
pressure
level (
Lev. Met.
).
The Sound Level Meter (SLM)
function
meets the standard IEC 61672-1:2013 for Class 2 accuracy.
You may also use 1/1 and 1/3-octave band real time analysis, dose meter (dosimeter) and reverberation
time options. These options broaden the main Level Meter functionality of the instrument, because 1/1
and 1/3-octave analysis as well as dose meter measurements are performed along with all calculations
of the broadband Level Meter results.
To activate a measurement function, open the
Meas. Funct
list and select with the
▲
/
▼
keys the
required function:
Level Meter
,
1/1 Octave
,
1/3 Octave
,
Dosimeter
or
RT60
.
<ENT>
Note: Type of measurement function is not displayed on the screen, so the user should
remember about the currently selected function!
Note: The 1/1 Octave, 1/3 Octave and RT60 functions are optional and should be unlocked
by entering an activation code in the text editor screen, which is opened after first attempt to
select it. Once unlocked, this function will be available permanently.
Optional functions that broaden the applications of the instrument can be easily installed. These optional
functions can be provided initially by the manufacturer or can be purchased later when required.
Note: It is not possible to change the measurement function during a measurement run. In this
case, the instrument displays for about 3 seconds the text:
“Measurement in Progress”. To
change the function of the instrument the current measurement must be stopped!
3.2
I
NSTRUMENT
’
S CALIBRATION
–
C
ALIBRATION
The instrument is factory calibrated with the supplied microphone for the reference environmental
conditions (see Appendix C). The microphone sensitivity is a function of the temperature, ambient
pressure and humidity, and when the absolute sound pressure level value is required, the absolute
calibration of the measurement channel should be performed. To select the calibration function, open the
Calibration
list.
Due to automatic calibration option the instrument can perform the sound calibration automatically, when
the calibrator is placed over the microphone (switched on or with the auto run function). The calibrator
signal is automatically detected, and the calibration measurement is started automatically. Just press
<Enter>
to confirm the calibration results. A sound measurement cannot be in progress while the
automatic calibration is being performed.
Note: SV 973 has two different dynamic ranges - one dedicated for the SLM functions (Level
Meter, 1/1 Octave, 1/3 Octave and RT60) and another one dedicated for the Dosimeter
function. Due to this the instrument uses two different calibration factors, which are both
calculated during the calibration.
Содержание SV 973
Страница 157: ...SVAN 973 User Manual Appendixes 157 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 158: ...SVAN 973 User Manual Appendixes 158 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 159: ...SVAN 973 User Manual Appendixes 159 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 172: ...SVAN 973 User Manual Appendixes 172 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 173: ...SVAN 973 User Manual Appendixes 173 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 174: ...SVAN 973 User Manual Appendixes 174 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 185: ...SVAN 973 User Manual Appendixes 185 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...
Страница 192: ...SVAN 973 User Manual Appendixes 192 LF filter according to EPA 93 F105 02 104 Low Frequency Noise Control Regulations...