![Svantek SV 971A Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/svantek/sv-971a/sv-971a_user-manual_1409270035.webp)
SV 971A User Manual
35
Note: In case of the infinite integration period or the infinite repetition cycles the measurement
should be stopped manually with the <Stop> key
.
Day time limits
The
Day Time Limits
parameter defines the day and night time limits required
by the local standards. These limits are used for the calculation of the
Lden
function (see Appendix D for definition). Two options are available:
6H-18H
and
7H-19H
. By default, it is set to
6H-18H
.
Note: In the Simple interface mode, the Day Time Limits parameter is hidden, but the
instrument will use settings previously defined in the Advanced mode or default settings
(6H-18H)
.
Detector type
The
LEQ Integration
parameter defines the detector
type for calculation of the
Leq
,
Lden
,
LEPd
,
Ln
and
Sel
results. Two options are available:
Exponential
and
Linear
. The formulae used for the
Leq
calculation are
given in Appendix D. Its value by default is set to
Linear
.
►
Linear
is required for obtaining the true RMS value of the measured signal. When this option is selected
values of the
Leq
,
Lden
,
LEPd
,
Ln
and
Sel
results do not depend on the detector time constant:
Fast
,
Slow
or
Impulse
(results are displayed without indication of detectors selected in profiles). In this case,
the indicator
Lin.
(or
L
) is displayed in different views.
Exponential
enables fulfilling the requirements of another standard for time averaged
Leq
measurements. When this option is selected values of the
Leq, Lden, LEPd, Ln
and
Sel
results depend
on the detector time constant. Results are displayed with the indicator of the detector type selected in
the profiles (
path: <Menu> / Measurement / Profiles
).
Note: In the Simple interface mode, the LEQ Integration
parameter is hidden, but the
instrument will use settings previously defined in the Advanced mode or default settings
(Linear)
.
Rolling Leq
In the two
LRx Time
positions, you can define integration periods for calculating
the
LR1
and
LR2
results (see Appendix D). Default values respectively:
30 m
and
60 m
.
4.2 S
ETTING MEASUREMENT TRIGGER
–
M
EASUREMENT
T
RIGGER
The
Meas. Trig.
position appears only in the
Advanced
interface mode and enables setting parameters of the
measurement trigger. The
Meas. Trig.
is a contexts list
of parameters in which the trigger (
Trigger
) can be
switched
Off
or On by selecting the trigger type (
Slope+
,
Slope
–
,
Level+
,
Level
–
or
G
). In case the
trigger is on, additional parameters can be defined: the
measurement result that is checked for a trigger
condition (
Source
), its threshold level (
Level
) and the
speed of the Source value changing (
Gradient
).
<ENT>
Note: In the Simple interface mode, the Measurement Trigger position is hidden, but the
instrument will use settings previously defined in the Advanced mode or default settings
(Trigger: Off)
.
The trigger condition is checked every 0.5 milliseconds.
Содержание SV 971A
Страница 220: ...SV 971A User Manual Appendixes 220 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 221: ...SV 971A User Manual Appendixes 221 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 222: ...SV 971A User Manual Appendixes 222 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 223: ...SV 971A User Manual Appendixes 223 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 224: ...SV 971A User Manual Appendixes 224 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Страница 232: ...SV 971A User Manual Appendixes 232 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 233: ...SV 971A User Manual Appendixes 233 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 234: ...SV 971A User Manual Appendixes 234 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 235: ...SV 971A User Manual Appendixes 235 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 236: ...SV 971A User Manual Appendixes 236 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Страница 245: ...SV 971A User Manual Appendixes 245 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter...
Страница 246: ...SV 971A User Manual Appendixes 246 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...
Страница 247: ...SV 971A User Manual Appendixes 247 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter...
Страница 254: ...SV 971A User Manual Appendixes 254 Z filter Class 1 according to IEC 61672 1 2013 Z filter characteristics...
Страница 273: ...SV 971A User Manual Appendixes 273 250 Hz 315 Hz 0 deg 90 deg 0 deg 90 deg 400 Hz 500 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 274: ...SV 971A User Manual Appendixes 274 630 Hz 800 Hz 0 deg 90 deg 0 deg 90 deg 1000 Hz 1250 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 275: ...SV 971A User Manual Appendixes 275 1600 Hz 2000 Hz 0 deg 90 deg 0 deg 90 deg 2240 Hz 2500 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 276: ...SV 971A User Manual Appendixes 276 2800 Hz 3150 Hz 0 deg 90 deg 0 deg 90 deg 3550 Hz 4000 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 277: ...SV 971A User Manual Appendixes 277 4500 Hz 5000 Hz 0 deg 90 deg 0 deg 90 deg 5600 Hz 6300 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 278: ...SV 971A User Manual Appendixes 278 7100 Hz 8000 Hz 0 deg 90 deg 0 deg 90 deg 8500 Hz 9000 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 298: ...SV 971A User Manual Appendixes 298 O ring gasket locations inside...