SV 971A User Manual
112
You can replace the microphone cartridge just unscrewing
the previous microphone and screwing the new one. When
unscrewing the microphone, care must be taken not to drop
the microphone. To avoid this, unscrew the microphone so
that it ended up in the user's palm.
Note: After replacing the microphone you must calibrate the instrument! Due to the significant
difference in the sensitivity of the microphones, the calibration drift may also be significant,
which in this case will not mean a malfunction.
The instrument is delivered with the attached preamplifier.
Nevertheless, sometimes it is necessary to disconnect the
preamplifier (for example, in the case SV 971A should be
used with the SV 271 monitoring station).
To disconnect the preamplifier from the instrument, unscrew
the screw threaded ring of the preamplifier and pull the
preamplifier out of the instrument.
To connect the preamplifier (with the microphone) to the
instrument, position the instrument socket and the
preamplifier plug in the way that the red point on the
instrument socket is in line with the pilot slot on the
preamplifier plug. Then insert the preamplifier plug into the
socket and tighten the screw threaded ring.
16.4 R
ESETTING THE INSTRUMENT
•
SYSTEM RESET:
internal software reset clears any setup configuration and brings back the
default factory settings. See
Factory Settings
(
path: <Menu> / Auxiliary Setup
).
•
HARDWARE RESET:
internal hardware reset doesn’t change any settings. Make sure the battery
is not exhausted, and the unit is turned off. Hold down the
<Shift>
and
<Start/Stop>
keys for
10 seconds, and then release them. Turn on the instrument as usually.
Note: Hardware reset should only to be used in extreme situations such as an instrument
hang-up.
Be aware, that a hardware reset:
- will stop any pre-programmed auto-run modes,
- will stop measurement run!
16.5 U
PGRADING THE FIRMWARE
SVANTEK is committed to continuous innovation path of development, and as such reserves the right to
provide firmware enhancements based on user’s feedback.
To update the instrument firmware:
•
Unpack the provided firmware package (provided as a suitable compressed file).
•
Make sure the instrument is turned off.
•
Connect the SC 158 cable to the instrument and then to the PC.
•
Keeping pressed the
<Enter>
and
<ESC>
keys switch on the instrument - the following
message should appear on the
instrument’s screen: Bootstrap ver: 2.01 (or higher).
•
Wait for the
“<USB>” message on the instrument’s screen and run the
go-usb.bat
file in the
PC.
•
The changing number and final message "..... o.k." should appear on the PC screen.
•
Successful firmware update will be indicated by the message
“
Program loaded!
”
•
Switch off the instrument.
Содержание SV 971A
Страница 220: ...SV 971A User Manual Appendixes 220 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 221: ...SV 971A User Manual Appendixes 221 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 222: ...SV 971A User Manual Appendixes 222 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 223: ...SV 971A User Manual Appendixes 223 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 224: ...SV 971A User Manual Appendixes 224 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Страница 232: ...SV 971A User Manual Appendixes 232 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 233: ...SV 971A User Manual Appendixes 233 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 234: ...SV 971A User Manual Appendixes 234 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 235: ...SV 971A User Manual Appendixes 235 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 236: ...SV 971A User Manual Appendixes 236 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Страница 245: ...SV 971A User Manual Appendixes 245 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter...
Страница 246: ...SV 971A User Manual Appendixes 246 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...
Страница 247: ...SV 971A User Manual Appendixes 247 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter...
Страница 254: ...SV 971A User Manual Appendixes 254 Z filter Class 1 according to IEC 61672 1 2013 Z filter characteristics...
Страница 273: ...SV 971A User Manual Appendixes 273 250 Hz 315 Hz 0 deg 90 deg 0 deg 90 deg 400 Hz 500 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 274: ...SV 971A User Manual Appendixes 274 630 Hz 800 Hz 0 deg 90 deg 0 deg 90 deg 1000 Hz 1250 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 275: ...SV 971A User Manual Appendixes 275 1600 Hz 2000 Hz 0 deg 90 deg 0 deg 90 deg 2240 Hz 2500 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 276: ...SV 971A User Manual Appendixes 276 2800 Hz 3150 Hz 0 deg 90 deg 0 deg 90 deg 3550 Hz 4000 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 277: ...SV 971A User Manual Appendixes 277 4500 Hz 5000 Hz 0 deg 90 deg 0 deg 90 deg 5600 Hz 6300 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 278: ...SV 971A User Manual Appendixes 278 7100 Hz 8000 Hz 0 deg 90 deg 0 deg 90 deg 8500 Hz 9000 Hz 0 deg 90 deg 0 deg 90 deg...
Страница 298: ...SV 971A User Manual Appendixes 298 O ring gasket locations inside...