background image

 

SV 34B User Manual 

1. 

INTRODUCTION 

 

1.1. Calibration 

Calibration determines the ratio between the input signal value and the indication on the display 
of the measuring device (or measured result).  

In  principle,  there  are  more  or  less  large  systematic  deviations  between  the  displayed 
measured value and the true value of the measured signal for every measuring device. The 
task of calibration is to determine these systematic deviations. 

The simplest way to consider such systematic deviations is a correction of measurements by 
calibration factor obtained with the use of an acoustic calibrator. 

However, in many cases it is sufficient to determine that the systematic deviations are within 
certain limits. 

"Correct" value of the measured signal is set before calibration, and this value is compared 
with a value measured by the measuring device. 

Thus, calibration means establishing a relationship between the reference and measured value 
of  the  variable  for  the  corresponding  standard  to  be  able  to  take  this  ratio  into  account  in 
subsequent measurements as a correction factor (calibration factor). 

 

1.2. Accuracy of calibration 

Measuring equipment and measurement methods have deviations. The measured variable is 
affected  by  environmental  conditions  (temperature  and  humidity),  as  well  as  the  operator's 
actions. The displayed value of the measured variable will therefore usually deviate from the 
true value of the measured variable. 

It  is  recommended  to  check  the  SV 34B  every  2  years  with  periodic  testing  of  the  test 
equipment to ensure that the level values do not change, and the test results are reliable. 

It is important to carefully check who should perform such monitoring: 

- internally by the monitoring body, 

- SVANTEK's own calibration laboratory according to ILAC, 

- externally from PTB = Physikalisch-Technische Bundesanstalt or 

- another local accredited laboratory. 

 

Accuracy 

There  is  a  deviation  between  the  true  value  and  the  average  value  of  the  series  of 
measurements under repetitive conditions, which is the result of repeated measurement of the 
reference level. 

 

Overview of the classification of sound level meters and calibrators 

Classification of sound level meters and acoustic calibrators 

The acoustic calibrators (see IEC 60942. 2003) and the sound  level meters (see IEC 61672: 
2002) are classified into their respective classes and types for accuracy. 

Содержание SV 34B

Страница 1: ...USER MANUAL SV 34B ACOUSTIC CALIBRATOR Warsaw 2019 03 20 Rev 1 01 Copyright 2019 SVANTEK All rights reserved...

Страница 2: ...3 1 1 Calibration 3 1 2 Accuracy of calibration 3 2 Acoustic calibrator SV 34B 5 2 1 General description 5 2 2 Using the calibrator 6 2 2 1 Button functions 6 2 2 2 Indicators 7 2 3 Replacing the bat...

Страница 3: ...libration factor 1 2 Accuracy of calibration Measuring equipment and measurement methods have deviations The measured variable is affected by environmental conditions temperature and humidity as well...

Страница 4: ...f 1kHz Class Type LS 1 2 Sound level meters dB 0 7 1 0 Acoustic calibrators dB 0 10 0 25 0 40 As shown in Table 1 the acoustic calibrator has unambiguously lower tolerances than the sound level meter...

Страница 5: ...f 1 kHz frequency is digitally generated and feeds the loudspeaker The sampled signal from the reference piezoresistive sensor indicates the level of currently generated signal in a feedback loop Base...

Страница 6: ...l cause full reset of the system Normally this function is not necessary It has been implemented in the case of inappropriate operation of the calibrator caused by external EM radiation subnormal atmo...

Страница 7: ...continuous light Note Calibration should not be performed until the 114 dB diode light is continuous The diode called LOW BAT presents the batteries status When the combined battery voltage is less t...

Страница 8: ...charged batteries d put two new batteries in place of the discharged ones with polarization as indicated on the printed board and calibrator s case e carefully put on the cover so that the diodes fit...

Страница 9: ...c max 10 sec Microphone dimensions 13 2 mm or with reduction adapter SA 30 Storage temperature range 25 C to 70 C 13 F to 158 F CE classification EN 61010 1 2010 EN 61326 1 2013 EN 60942 2003 Working...

Страница 10: ...RF emission in the direction parallel to the axis of the calibrated microphone in the acoustic chamber The lowest level of noise immunity is parallel to the axis of the calibrated microphone in the a...

Отзывы: