SV 200A User Manual
4.3 C
ALIBRATION WITH THE USE OF THE CONTROL PANEL
The calibration via the control panel (manual calibration) gives the user an option to decide whether the new
calibration factor should replace the current one.
Note: Before proceeding with the manual calibration you should be sure that the Automatic
calibration is Off (see Chapter
To perform the manual calibration, follow next steps:
1. Uncover the control panel of SV 200A
by unlocking
and shifting the control panel flap.
2. Press the
and
keys simultaneously to open the Menu
window of the panel display
.
3. Press
the
key to open the
Function
position, select the
Calibration
position with the
key and press
the
key to
open it.
4. Press
the
key to open the
By Measurement
position and set
up the required calibration
Level
according to the calibration card of
your calibrator. Press
the
key to confirm the new calibration
level.
5. Attach the calibrator on the microphone and switch it on (if the used
calibrator doesn’t have switch-on automatic feature).
6. Start the calibration measurement by pressing the
key. During
the calibration measurement, the level of the calibration signal will
be displayed. If three consecutive 1-second results are stable within
±0.1dB
margin, the calibration measurement is stopped, and the
calibration factor is calculated. Otherwise the instrument will stop the
measurement and display the message “Failed!” after 10 seconds
from the calibration measurement start.
7. The successful calibration will result in calculation of the new
calibration
Factor
, which should be confirmed (
Yes
) with the
key or rejected (
No
) with the
key. After confirmation, the
calculated calibration factor will be the current calibration factor.
8. If the calculated calibration factor is out of the range
±20dB
, the
calibration will fail and the message “Failed!” will appear on the
display.
9. Detach the calibrator from the microphone.
Содержание SV 200A
Страница 123: ...123 SV 200A User Manual Appendix A REMOTE CONTROL firmware revision 1 x x Under development ...
Страница 153: ...153 SV 200A User Manual Free Field Frequency response of SV 200A ...
Страница 164: ...164 SV 200A User Manual 250 Hz 315 Hz 90 deg 0 deg 90 deg 0 deg 400 Hz 500 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 165: ...165 SV 200A User Manual 630 Hz 800 Hz 90 deg 0 deg 90 deg 0 deg 1000 Hz 1250 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 166: ...166 SV 200A User Manual 1600 Hz 2000 Hz 90 deg 0 deg 90 deg 0 deg 2240 Hz 2500 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 167: ...167 SV 200A User Manual 2800 Hz 3150 Hz 90 deg 0 deg 90 deg 0 deg 3550 Hz 4000 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 168: ...168 SV 200A User Manual 4500 Hz 5000 Hz 90 deg 0 deg 90 deg 0 deg 5600 Hz 6300 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 169: ...169 SV 200A User Manual 7100 Hz 8000 Hz 90 deg 0 deg 90 deg 0 deg 8500 Hz 9000 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 170: ...170 SV 200A User Manual 9500 Hz 10000 Hz 90 deg 0 deg 90 deg 0 deg 10600 Hz 11200 Hz 90 deg 0 deg 90 deg 0 deg ...
Страница 177: ...177 SV 200A User Manual 8 0 kHz 1 1 octave filter 4 0 kHz 1 1 octave filter 2 0 kHz 1 1 octave filter ...
Страница 178: ...178 SV 200A User Manual 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter ...
Страница 179: ...179 SV 200A User Manual 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter ...