47
SVAN 977C User Manual
Weighting filter
Sound measurements
•
Z
according to IEC 61672-1:2013 for Class 1,
•
A
according to IEC 651 and IEC 61672-1:2013 for Class 1,
•
C
according to IEC 651 and IEC 61672-1:2013 for Class 1,
•
B
according to IEC 651 for Class 1,
•
U
according to IEC 61012:1990 for Class 1,
•
AU
according to IEC 61012:1990 for Class 1,
•
LF
low frequency (20-200 Hz) Real-Time Band Pass Filter.
Vibration measurements
•
acceleration measurements:
HP
,
HP1
,
HP3
,
HP10
,
Wh
,
•
velocity measurements:
Vel1
,
Vel3
,
Vel10
and
VelMF
,
•
displacement measurements:
Dil1
,
Dil3
and
Dil10
.
Characteristics of these filters are given in Appendix C.
RMS detector
Available RMS detectors (time constants):
Impulse
,
Fast
and
Slow
for
Sound
measurements and
100ms
,
125ms
,
200ms
,
500ms
,
1.0s
,
2.0s
,
5.0s
,
10.0s
for Vibration
measurements.
You can easily access the
Profiles
screen
from the result view screen. Select some
profile field (for example,
P(1)
) with the ▲ / ▼
or
◄ / ► key and press
<ENTER>
.
<ENT>
4.4 C
ONFIGURING DATA LOGGING
–
L
OGGING
Summary Results
(
Peak
,
Max
,
Min
,
Spl
,
Leq
,
SEL
,
Lden
,
LEPd
,
Ltm3
,
LTeq
, 10 x
Ln
,
EX
,
SD
,
LR1
,
LR2
,
OVL
,
NC
,
NR
for Sound measurements or
RMS
,
Ovl
,
Peak
,
P
–P
for Vibration measurements as well as
1/1-octave, 1/3-octave or FFT spectra with Total results and meteo results) are measured and saved in the file
with the step defined by the
Integration Period
parameter as many times as defined by the
Repetition Cycles
parameter (
path: <Menu> / Measurement / General Settings
).
The instrument enables also additional registration of some results with different step defined by the
Logger Step
parameter (
path: <Menu> / Measurement / Logging / Logger Setup
). Therefore, it is possible to save in parallel
two sequences of measured results
– one for
Summary Results
and another for so called
Logger Results
.
When logging is enabled, selected logger results from three independent profiles (
Peak
,
Max
,
Min
,
Leq
,
LXY
,
LR1
and
LR2
for Sound measurements or
Peak
,
P-P
,
Max
and
RMS
for Vibration measurements as well as 1/1-
octave or 1/3-octave spectra with Total results and meteo results) can be saved simultaneously with time step
down to
2ms
. Recording of logger results to a file is stopped after the period, which is equal to
Integration Period
multiplied by
Repetition Cycles
or after stopping a measurement manually.
Note: LXY is an abbreviation for the L (or Spl) result measured with X filter and Y detector time
constant
– see Appendix D.
Содержание Shawcity SVAN 977C
Страница 231: ...231 SVAN 977C User Manual Appendixes 630 Hz 800 Hz 1000 Hz 1250 Hz 1600 Hz 2000 Hz...
Страница 232: ...232 SVAN 977C User Manual Appendixes 2240 Hz 2500 Hz 2800 Hz 3150 Hz 3550 Hz 4000 Hz...
Страница 233: ...233 SVAN 977C User Manual Appendixes 4500 Hz 5000 Hz 5600 Hz 6300 Hz 7100 Hz 8000 Hz...
Страница 234: ...234 SVAN 977C User Manual Appendixes 8500 Hz 9000 Hz 9500 Hz 10000 Hz 10600 Hz 11200 Hz...
Страница 242: ...242 SVAN 977C User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 243: ...243 SVAN 977C User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 244: ...244 SVAN 977C User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 245: ...245 SVAN 977C User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 246: ...246 SVAN 977C User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Страница 250: ...250 SVAN 977C User Manual Appendixes...
Страница 262: ...262 SVAN 977C User Manual Appendixes 2 0 kHz 1 1 octave filter 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter...
Страница 263: ...263 SVAN 977C User Manual Appendixes 250 Hz 1 1 octave filter 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter...
Страница 264: ...264 SVAN 977C User Manual Appendixes 31 5 Hz 1 1 octave filter 16 0 Hz 1 1 octave filter 8 00 Hz 1 1 octave filter...
Страница 265: ...265 SVAN 977C User Manual Appendixes 4 00 Hz 1 1 octave filter 2 00 Hz 1 1 octave filter 1 00 Hz 1 1 octave filter...
Страница 311: ...311 SVAN 977C User Manual Appendixes 250 Hz 315 Hz 90 deg 0 deg 90 deg 0 deg 400 Hz 500 Hz 90 deg 0 deg 90 deg 0 deg...
Страница 312: ...312 SVAN 977C User Manual Appendixes 630 Hz 800 Hz 90 deg 0 deg 90 deg 0 deg 1000 Hz 1250 Hz 90 deg 0 deg 90 deg 0 deg...
Страница 313: ...313 SVAN 977C User Manual Appendixes 1600 Hz 2000 Hz 90 deg 0 deg 90 deg 0 deg 2240 Hz 2500 Hz 90 deg 0 deg 90 deg 0 deg...
Страница 314: ...314 SVAN 977C User Manual Appendixes 2800 Hz 3150 Hz 90 deg 0 deg 90 deg 0 deg 3550 Hz 4000 Hz 90 deg 0 deg 90 deg 0 deg...
Страница 315: ...315 SVAN 977C User Manual Appendixes 4500 Hz 5000 Hz 90 deg 0 deg 90 deg 0 deg 5600 Hz 6300 Hz 90 deg 0 deg 90 deg 0 deg...
Страница 316: ...316 SVAN 977C User Manual Appendixes 7100 Hz 8000 Hz 90 deg 0 deg 90 deg 0 deg 8500 Hz 9000 Hz 90 deg 0 deg 90 deg 0 deg...
Страница 336: ...336 SVAN 977C User Manual Appendixes O ring gasket locations inside...