Remove the six (6) shipping screws (indicated by the red stripes)
IMPORTANT: Retain the six (6) shipping screws and keep them with the instrument
packaging in the event that the instrument needs to be shipped to another location.
6.
Carefully place the manipulator on the base plate so that the micropipette holder clamp
faces toward the front of the microscope, as shown above for a right-hand
micromanipulator. The two innermost screw holes (closest to the microscope – arrowed)
on the base of the mechanical should be visible. Attach the manipulator to the adapter
plate in the position shown, using the screws and tool provided. Do not use shipping
screws for this purpose - they are too long.
7.
Slide the dovetail on the bottom of the adapter plate into the slide of the microscope
adapter (it may be necessary to loosen the locknut on the rear of the microscope adapter).
Do not tighten the locknut at this time.
Figure 2-2. Engaging the manipulator base plate in the microscope adaptor.
8.
Unpack the controller cabinet and place it on a shelf or flat surface near the microscope.
9.
Power and Voltage Selection: The power entry module on the rear of the controller has a
voltage-select dial that can be set to either “110” (100 – 120) or “220” (200 – 240) VAC.
Ensure that the voltage rating matches your local voltage supply; if not, switch it now.
10.
Power Fuse, Ratings, and Selection: The power entry module also contains the fuse. It is
very important that the correct fuse rating is used for the voltage selected, as indicated in
the table below.