4.10 20141208
(
)
9
2.
INSTALLATION
The directions below describe the installation procedure common to both left- and right-
handed micromanipulators, although the figures illustrate a right-hand micromanipulator (a
left hand micromanipulator would be illustrated as a mirror image of these figures). It is
extremely important that these installation instructions are followed closely.
1.
Unpack and assemble the microscope adapter and attach it firmly to the microscope
according to the instructions included with the adapter.
2.
Unpack the mechanical manipulator adapter base plate. The Y-axis slide on the bottom of
the adapter plate can be adjusted so that the mechanical is positioned correctly with
respect to the microscope’s optical axis. Each model microscope will require special Y-axis
slide settings. To determine the appropriate setting for your microscope, refer to the
lookup table contained in the instruction sheet supplied with the microscope adapter.
3.
Align the back edge of the slide in the numbered channel with the number in the
instructions that corresponds with the microscope being used. Lock the Y-axis slide
screws firmly with the hex (Allen) key provided.
Figure 2-1. Mounting the right-handed manipulator to the far left corner of the adapter plate.
4.
Unpack the manipulator mechanical. Take extra care while handling the mechanical -
before it is mounted on the microscope adapter, it is vulnerable to impacts which can
damage the stepper motors.
XENOWORKS MICROMANIPULATOR SYSTEM (TOUCHDECLUTCH) OPERATION MANUAL – REV.