P2000/G Laser-Based Micropipette Puller Quick Start Guide 

Warning  to  Users: 

Never  adjust  the  black  micropositioner  which  holds  the  retroreflective  mirror  (retromirror)  without  explicit 

instruction from Sutter Instrument Co. This has been carefully calibrated to ensure optimal performance of your instrument.

 

First Time Setup 

1.

 

Remove  the  instrument  from  the  box  and  place  on  a  flat  surface.  Retain  the  box  and  packing 
materials. These are needed in case the instrument needs to be shipped. 

2.

 

Lift the acrylic lid and remove the two red plastic rectangles. Retain these in case the instrument 
needs to be returned to the factory. Remove the rubber bands holding the puller bars together. 

3.

 

Loosen the shipping screw by turning it counterclockwise several turns. Leave the screw partially 
threaded into the puller. You will need to retighten this screw if the instrument needs to be shipped.  

4.

 

Insert the supplied power cord in the power entry module on the back of the instrument. Plug the 
cord into a power source. 

5.

 

Power the unit on by toggling the switch on the left side of the puller. 

6.

 

Ensure  the  tilt  setting  has  not  changed  during  shipping  by  running  a  thermal  paper  test.  The 
following sections describe accessing a program and loading glass, which must be done prior to this 
test. 

 

Selecting and Loading a Program 

1.

 

At power on, the home screen prompts you to select a program. Using the keypad, enter a number from 0-99 and press <ENTR> 
to access that program. Some programs were loaded at the factory. Refer to the operations manual for detail on these. 

2.

 

Each program has 5 parameter spaces (HEAT, FIL, VEL, DEL, PUL). These spaces can be populated with values in either of 2 ways: 

i.

 

Entering 3 digits will fill a parameter space and advance to the next. For values less than 100, values can be loaded with 
leading zeros (e.g. 005, 040, etc.) 

ii.

 

Entering fewer than 3 digits followed by the <ENTR> key. You can advance to the next parameter without changing the 
value of the current one by keying <ENTR> without entering any digits. 

3.

 

Exit a program and return to the home screen by depressing the black <RESET> switch 

Loading Glass 

1.

 

Loosen the glass clamps on both puller bars by turning the knobs counterclockwise. 

2.

 

Insert a piece of capillary glass into the 

top

 groove of one puller bar. Start with the 

whole capillary behind the glass clamp. 

3.

 

With  your  thumb  near  the  end  of  the  capillary  farthest  from  the  shroud,  slide  the 
capillary towards the center of the puller until the far end of the capillary is outside of 
the cutout in the puller bar. 

4.

 

Still holding the back end of the capillary, clamp the glass snugly. 

5.

 

Release the puller bars from their clip locks by depressing the flat metal spring behind 
each bar and then pulling the bars forward slightly. 

Tilt micrometer with 
shipping screw partially 
removed 

 

Load glass here 

Do NOT load glass here 

Step 3 