15
(numerical values may vary). Please refer to the FRONT PANEL chapter and the beginning
of this chapter for a full description of the PARAMETERS shown.
UNLOCKED P=500 JAN 18, 1997 2:30 PM
3 1 HEAT=525,PULL=150,VEL.= 75,TIME=250
Figure 3-3. P-97 Program display.
To select a different program,
Press < RESET > to bring up the sign-on prompt then
press the number of the desired program (0 to 99)
followed by < ENTR >.
3.2.6
Viewing a Program (< NEXT > / < LAST >)
The front panel display limits you to viewing only two adjacent lines of a PROGRAM at one
time. When you first select a PROGRAM, the display above will appear. The top line of every
program is a header line, which displays the PRESSURE setting, the WRITE PROTECT
STATUS and the LAST EDIT DATE. The second line of the display is Line 1 of program.
To view additional lines of the
program
Press the < NEXT > key
To scroll to previous lines
Press the < LAST > key.
3.2.7
Clearing a Program from Memory (< CLR >)
To clear the values of a particular program from memory you can utilize the CLEAR
function.
Pressing the < CLR > key gives you the following message:
DO YOU WISH TO CLEAR ALL VALUES FROM THE
PRESENT CYCLE TO THE END? NO=0 YES=1
Figure 3-4. Clear Program Display.
Entering < 1 > will clear the PROGRAM from the line the cursor is on to the end of the
PROGRAM. If the cursor is on Line 1, the entire program will be cleared from memory. If the
cursor is on Line 2, Line 1 will be preserved, and lines 2-8 will be cleared.
3.2.8
Editing a Program
Select a program number (e.g., 10). If there are already values entered, make sure that this
program was not entered by another user of the puller. Unused program areas are usually
cleared before a puller is shipped, but occasionally test program values are inadvertently left
in memory. We recommend that unused programs be cleared before proceeding.
P-97 FLAMING/BROWN MICROPIPETTE PULLER OPERATION MANUAL – REV. 2.43 - DOM (20161118)
Содержание P-97
Страница 3: ......
Страница 4: ......
Страница 48: ...P 97 FLAMING BROWN MICROPIPETTE PULLER OPERATION MANUAL REV 2 43 DOM 20161118 38 This page intentionally blank...
Страница 54: ...P 97 FLAMING BROWN MICROPIPETTE PULLER OPERATION MANUAL REV 2 43 DOM 20161118 44 This page intentionally blank...
Страница 56: ...P 97 FLAMING BROWN MICROPIPETTE PULLER OPERATION MANUAL REV 2 43 DOM 20161118 46 This page intentionally blank...
Страница 60: ...P 97 FLAMING BROWN MICROPIPETTE PULLER OPERATION MANUAL REV 2 43 DOM 20161118 50 This page intentionally blank...
Страница 66: ...56 NOTES P 97 FLAMING BROWN MICROPIPETTE PULLER OPERATION MANUAL REV 2 43 DOM 20161118...