background image

 

LAMBDA 421 OPERATION MANUAL – REV. 1.00N  (20170623) 

10 

still further reduce the current proportionally so their operation will be appear to be the 
same as the other channels even though the range of currents available has been reduced.  

3.7

 

Control via the DG-4 Interface 

In order to understand the Lambda 421 it is important to know the way the DG-4 functions. 
The DG-4 selects one of 4 filter paths for the light source. The filter values, in the DG-4 
context, are actually just memory locations for the settings assigned to a filter value for 
positioning the two galvanometers in the optical system. These settings could be any value in 
the range of 0 to 4095. There are 16 available filter values in the DG-4 command set. Filter 
values can be any number from 0 to 15. Filter 0 is normally reserved as a shutter function 
and may have 0 stored for both galvanometers. Filters 1, 2, 3, and 4 typically have settings 
that deliver the light through one of the four physical optical filters to the output port at full 
power. Filters 5-12 are typically programmed to select one of the four filter paths at reduced 
power. The standard setup uses filters 5-8 to select the filters used in 1-4 at 50% output 
power and 9-12 are 1-4 at 33% output power. Filters 13-15 are available but normally defined 
at the factory.   
We have produced the same sort of control for four LEDs using the new USB DG-4 board. 
This approach produces an LED version of the DG-4 that will run with existing software that 
supports the DG-4. Since the DG-4 only works with one filter active at a time this command 
set does not allow computer selection of more than 1 LED at a time. We have added some 
basic logic circuits that convert the value of the filter selected to a signal that activates a 
specific LED driver.  Each LED is gated by a logic level signal and all LEDs will have the 
analog input connected to the common DAC output on the DG-4 circuit board.  
Since only one LED is activated by the DG-4 control at a time we can use a single DAC to 
supply the current-controlling analog voltage for all four LEDs. In the DG-4 this DAC is 
normally used for setting a galvanometer command signal. The DAC value is automatically 
set to a value associated with a given filter when that filter is selected. We simply connected 
that DAC output to the analog current control inputs of all 4 LED drivers and we now have a 
system that automatically sets the desired current for each LED as it is selected. The value 
for the DAC can be stored and read back using simple software commands.  
The logic level signals that active the correct LED for each filter value are derived from the 4 
bit binary output normally logic level output used by the DG-4 to indicate the selected filter. 
We have added logic that converts the binary filter value into four separate TTL level 
outputs, one for each LED. The original binary output is also available as before. The logic 
was designed so that filter values 1-4 select the corresponding LEDs, 1-4. Filter values 5-8 
and 9-12 also select LEDs 1-4. Filter values 13-15 select LEDs 1-3. Filter value 0 turns off all 
LEDs. Looking at this in another way: 
LED 1 is activated by filter values 1, 5, 9   and 13.  
LED 2 is activated by filter values 2, 6, 10 and 14.  
LED 3 is activated by filter values 3, 7, 11 and 15.  
LED 4 is activated by filter values 4, 8, 12  
Filter 0 does not active any LED.  
The DG-4 controller stores two DAC values for each of the 16 possible filter values. We have 
connected the analog output of one of these two DACs to the analog current control input for 

Содержание Lambda 421

Страница 1: ...28 Web www sutter com Fax 415 883 0572 Email info sutter com Lambda 421 Ultra High Speed Optical Beam Combined 4 Channel Wavelength Switching LED Light Source System With Lambda DG 4 External Control...

Страница 2: ...II Copyright 2017 Sutter Instrument Company All Rights Reserved...

Страница 3: ......

Страница 4: ...2...

Страница 5: ...0 Hz line voltage This instrument is designed for use in a laboratory environment that has low electrical noise and mechanical vibration Surge suppression is recommended at all times Fuse Replacement...

Страница 6: ...is instrument is not recommended and if undertaken is done so at the users own risk Do not operate if there is any obvious damage to any part of the instrument Avoiding Physical Injury while Powered u...

Страница 7: ...1 3 Internal External Control Switch INT EXT 7 3 1 4 Dimness Intensity Control Knob 0 11 7 3 2 Indicators 8 3 2 1 Power Light 8 3 3 Connectors 8 3 3 1 Light Beam Combiner Output 8 3 3 2 STROBE BNC Co...

Страница 8: ...s controls and connectors 2 Figure 1 2 Stages of pass through and reflection from each input light source 3 Figure 1 3 Lambda 421 combined output based on the pass through and reflection of each input...

Страница 9: ...of the system leaving more in depth technical information towards the end of the chapter The next chapter describes how to install the system followed by one or more chapters that provide operation in...

Страница 10: ...urces In the diagram below the optical paths are outlined for each position including the reflections that occur Traditionally combining more than two light sources required the use of a dichroic ladd...

Страница 11: ...er losses than previously achievable As an added benefit the last position in the optical train does not require any filter since no other input reflects from that position This input can be used with...

Страница 12: ...toward shorter wavelengths While it would be ideal to have a coating optimized for this application we have found that stock STR filters can be used if you correct for the shift in the band pass when...

Страница 13: ...ight Guide Liquid Light Guide dovetail with C mount extension Small Zeiss female dovetail USB Interface Cable 9 pin Serial Interface Cable 25 pin Parallel Interface Cable Manual One or more additional...

Страница 14: ...6 WARNING DO NOT LOOK DIRECTLY INTO THE LIGHT GUIDE The output of the light or the light guide should be directed into the microscope using the appropriate adapters directed away from anyone s eyes a...

Страница 15: ...annel Does not affect externally supplied light source TTL Internally supplied LED light source on off state is controlled by TTL triggering via STROBE 3 1 3 Internal External Control Switch INT EXT T...

Страница 16: ...th the SERIAL interface The USB interface is enabled at the factory when the Lambda 421 is shipped unless the SERIAL interface is specifically requested When the USB interface is enabled the SERIAL in...

Страница 17: ...o the top position ON turns the on When selecting the lowest position TTL the channel is controlled by the DG 4 interface circuit board The switch to the left is labeled INT EXT In the upper position...

Страница 18: ...its that convert the value of the filter selected to a signal that activates a specific LED driver Each LED is gated by a logic level signal and all LEDs will have the analog input connected to the co...

Страница 19: ...of any channel if desired The control of the LEDs then is entirely from the DG 4 interface The connections for the DG 4 interface are all available on the back panel Figure 3 2 Rear panel of the Lamb...

Страница 20: ...ER VALUE LED SELECTED SUGGESTED CURRENT 0 NONE 0 1 1 100 2 2 100 3 3 100 4 4 100 5 1 50 6 2 50 7 3 50 8 4 50 9 1 33 10 2 33 11 3 33 12 4 33 13 1 15 14 2 15 15 3 15 The actual attenuation is determined...

Страница 21: ...ne Maintenance Routine cleaning of the Lambda 421 system is required to prevent excessive dust accumulations Wipe all exterior surfaces with a dry soft cotton cloth Periodically inspect all cables ens...

Страница 22: ...LAMBDA 421 OPERATION MANUAL REV 1 00N 20170623 14 This page intentionally blank...

Страница 23: ...d to each of the 4 LED channels is adjustable The LED s vary in maximum current from 1A to 5A and adjustment process is straightforward You will the following tools Multimeter 1 5mm Allen wrench Small...

Страница 24: ...an LED rated for 5A and switching to an LED rated for 1 or 2A there is a real risk that you can damage the LED accidentally Please follow these instructions carefully 5 2 Adjustment Procedure for Swit...

Страница 25: ...5mm allen wrench on lens tube 3 Remove and replace the Filter assembly 4 Turn your Multimeter ON and set to Volts DC 5 Turn on the Main power of the Lambda 421 6 With the LED connected to channel tha...

Страница 26: ...LAMBDA 421 OPERATION MANUAL REV 1 00N 20170623 18 FG OBC 561 2 FG OBC 590 2 FG OBC 630 5 FG OBC 660 2 FG OBC 740 1 FG OBC 850 1 FG OBC 940 1...

Страница 27: ...id this warranty Warranty work will be performed only at the factory The cost of shipment both ways is paid for by Sutter Instrument during the first three months this warranty is in effect after whic...

Страница 28: ...LAMBDA 421 OPERATION MANUAL REV 1 00N 20170623 20 This page intentionally blank...

Страница 29: ...wer cord from the power entry module Press down on the lever that is located just above the fuse holder and below the power receptacle and pry the holder straight out of the power entry module Figure...

Страница 30: ...LAMBDA 421 OPERATION MANUAL REV 1 00N 20170623 22 This page intentionally blank...

Страница 31: ...Dimensions W x D x H 15 75 x 11 x 7 5 in 40 x 27 9 x 19 05 cm Weight 17 8 lbs 8 07 kg Light Guide Core Type Liquid Length 2 m 6 5 ft Diameter 3 mm 0 11 in Electrical Mains voltage 110 240 V 50 60 Hz P...

Страница 32: ...67 251 See Note 1 Connected to metal faceplates of connectors on both ends See Note 2 USB USB A USB B 10 ft 3 048 m Note 1 Fair Rite Products Corp P O Box J One Commercial Row Wallkill NY 12589 USA No...

Страница 33: ...D 00001101 0013 M 1 Command Move on command to LED 1 at factory defined level 14 0E 00001110 0014 N 2 Command Move on command to LED 2 at factory defined level 15 0F 00001111 0015 O 3 Command Move on...

Страница 34: ...and allowing both the before and after commands which have the same value to be accepted as valid commands Execute Next Instance of Previous Command Rx All 2 0 Echoed command sequence 1 13 0D 0000 110...

Страница 35: ...0 0000 0000 0000 1111 1111 0000 1111 0000 0000 0255 0015 Filter galvo DAC value encoded into a 16 bit unsigned short 2 bytes Little Endian Set Galvo A or Galvo B DAC Values for All Nominal Filter Posi...

Страница 36: ...l 2 0 Echoed command sequence 1 13 0D 0000 1101 CR Completion indicator Tx All 1 0 242 F2 1111 0010 0242 Returns to normal on line mode End Run from Ring Buffer Rx All 2 0 Echoed command sequence 1 13...

Страница 37: ...Buffer Status 1 1 4 01 04 0000 0001 0000 0100 Mode state 2 13 0D 0000 1101 CR Completion indicator Get Ring Buffer Contents Tx All 2 0 245 F5 1111 0101 0245 Get the contents of the ring buffer Rx All...

Страница 38: ...or Galvo B DAC Values for All Nominal Filter Positions 0 4095 0000 FF0F 0000 0000 0000 0000 1111 1111 0000 1111 0000 0000 0255 0015 Filter galvo DAC value encoded into a 16 bit unsigned short 2 bytes...

Страница 39: ...f two bytes ordered in Little Endian least significant byte last format see NOTE 3 The value stored in these two bytes is always unsigned meaning that the value will always be positive negative values...

Страница 40: ...LAMBDA 421 OPERATION MANUAL REV 1 00N 20170623 32 This page intentionally blank...

Страница 41: ...ctions 5 introduction 2 features 2 general description 2 technical support 1 L LED control switch 7 LED control switch 7 light beam combiner output 8 M mains fuses iii Maintenance 13 N notes user 34 O...

Страница 42: ...cations 23 dimensions 23 light guide 23 electrical 23 cables 23 mains fuse 23 mains voltage 23 power cord 23 lamp life 23 output range 23 lamp 23 light guide dimensions 23 weight 23 technical support...

Отзывы: