Additional commands for Analog output units
Command
Description
Command Syntax
(command underlined)
ACCUBUBBLE response
(underlined)
"a" represents the single-
character address
XAR Set
Analog
Output range
aXARzf!
Where z is the pressure in psi to be
represented by
0.000 VDC
f is pressure in psi to be
represented by 5.000 VDC
All input values are standard
SDI-12 values with polarity sign.
a0012
indicating the command will be
complete in one (1) second and that
two (2) values can be collected.
NOTE:
A D0 command issued after
the XAR will return two (2) values
representing the pressure range that the
user has selected.
XAO Set
Analog
Output to a fixed
voltage
aXAOv!
where v is the required output in
VDC
All input values are standard
SDI-12 values with polarity sign
a0011
indicating that the command will take
one (1) second to complete and that
one (1) value can be collected
NOTE:
A D0 command issued after
the XAO command will return the D/A
value corresponding to the requested
voltage.
The unit's output will not act as a
pressure sensor again until this same
command is given with a negative
voltage request.
XOM Set operating
mode
aXOMm!
where m is the desired operating
mode.
values for m
0
Bubbler Operation disabled
.
Quadrature output disabled.
(Low power consumption for
Accububbles being used as a
pressure sensor.)
64 Bubbler operation enabled.
80
Bubbler operation enabled.
SDI-12 measurements not
required for the analog output
to be updated.
NOTE:
When selecting mode 80,
also review the pump cycle time
set with the XPT command.
All input values are standard
SDI-12 values with polarity sign
attt1
indicating that the command will take
ttt seconds to complete and that one (1)
value can be collected
NOTE:
A D0 command issued after
the XOM command will return the set
operating mode.
42