Technical Data
EN
13.1
Characteristics of circuit breaker
Effect of ambient temperature
The units are calibrated for an ambient temperature of 23 °C. To determinate the
rated current for a lower or higher ambient temperature, use a correction factor
(typical value) from the table below:
Example: Rated current = 5 A; Enviroment temperature = 50 °C; --> Correction
factor = 1.04; Resulting current = 5.2 A --> Fount to next higher rated current: 6 A
2. See DIN EN 60
3. Acc. to 93/42/EEC
4. Formerly "repetition accuracy" - describes the reproducibility of a measurement ("internal precision")
5. Degree of match between displayed (actual) and correct (nominal) value; typical info "x% Full Scale"
Ambient temperature (°C)
Correction factor
0
0.90
+23
1.00
+40
1.03