32
Safety Functions
EN
7.3
Safety Functions in AirSeal Mode
Stop insufflation / System calibration
The pressure is reduced slowly before the system comes to a complete stop when
stopping AirSeal® insufflation.
When procedure is completed, AirSeal Mode will undergo a final calibration prior
to shut down. During this process no pressure or insufflation is built up.
Severed tube connection
If the connection of the AirSeal® filter tube set with the AirSeal® cannula is inter-
rupted or severed, the system stops and the warning
Check connection of Tri-Lu-
men Filtered Tube Set!
is depicted and a 15 second countdown begins. During
these 15 seconds, AirSeal functionality will automatically restart if the tri-lumen
tube set is reconnected to the AirSeal® cannula. Alternatively, insufflation func-
tionality will automatically restart if the Single Lumen Adapter is reconnected to
the tri-lumen tube set and to a conventional cannula.
Leakage warning
If a large leak makes it impossible to maintain the AirSeal® function, the warning
message
Check for excessive leakage and suction
is depicted.
7.4
Fill Level Display
Tri-lumen and bifurcated filtered tube sets are equipped with fluid sensors to
monitor the fluid level in the filters and to warn users of a possible device con-
tamination.
When working in the
AirSeal Mode
and
Smoke Evacuation Mode
, body fluids may
be flushed from the surgical field through the evacuation line and into the filter
housing. In order to prevent contamination of the device, bodily fluid is retained
by the fluid trap of the housing component. The capacity of the fluid trap is lim-
ited.
Fluid level LOW (AirSeal Mode and Smoke
Evacuation Mode)
A warning message and an audible signal will be emitted if the fill level
LOW
is
reached. Cannula and tubing placement should be checked to make sure no more
fluid enters the filter. Change the filtered tube set, and insufflation can be con-
tinued with full functionality.
Fluid level HIGH (Smoke Evacuation Mode)
If fluid level
HIGH
is reached in Smoke Evacuation Mode: Smoke Evacuation func-
tion will stop but insufflation can still be continued.
The iFS should be inspected to assess any possible damage. Call for service.
Fluid level HIGH (AirSeal Mode)
If fluid level
HIGH
is reached in AirSeal Mode, AirSeal function will stop. You will
be prompted to change to the tri-lumen tube set to resume AirSeal function. The
iFS should be inspected to assess any possible damage. Call for service.
Note:
If no new tri-lumen tube set is available, the iFS will continue with insuffla-
tion (one-way) with the existing tri-lumen tube set. This requires that the obtu-
rator is reinserted into the AirSeal® cannula or that the tri-lumen tube is
connected with the single lumen adapter to a conventional cannula after con-
ventional cannula is inserted. If AirSeal® cannula is removed, the surgeon may
wish to insert a conventional cannula in same entry point.
Содержание AirSeal iFS
Страница 1: ...EN Instructions for use...
Страница 4: ......
Страница 6: ...2 EN 15 AirSeal iFS Warranty 52 16 Test Log 53 16 1 Test Log 53 16 2 Return Form 54 Index 55...
Страница 20: ...16 Initial Device Startup EN House gas supply pressure OK House gas supply pressure too low...
Страница 57: ...Test Log 53 EN 16 Test Log 16 1 Test Log Date Results Comment Signature...
Страница 60: ...W55 186 EN AA...