Chapter 5 Configuration
www.supmea.com
16
RS485 Setting: set 485 communication address
(
0~255
)
and baud rate
(
2400
、
4800
、
9600
、
19200
)
.
Current Transmission: set 4mA corresponding value and 20mA
corresponding value of 4-20mA output.
5.5. Alarm setting
DM3000
-------------------- Alarm setting --------------------
1. DO High Alarm
2. DO Low Alarm
DO High Alarm: when the measured value is higher than high alarm pull-on
value, high alarm relay pulls on; when the measured value is lower than
high alarm cut-off value, high alarm relay cuts off.
DO Low Alarm: when the measured value is lower than low alarm pull-on
value, low alarm relay pulls on; when the measured value is higher than
lower alarm cut-off value, low alarm relay cuts off.
DM2800
-------------------- Alarm setting --------------------
1.
DO High Alarm
2.
DO Low Alarm
3.
SAT High Alarm
4.
SAT Low Alarm
5.
OPP High Alarm
6.
OPP Low Alarm