10.4. Transducer mounting methods
W210 insertion transducer mounting method: Z method through M24, it should be
installed according to the specific application condition.
Important
For horizontal pipelines, transducers must be fixed on the sides of the pipe (i.e. at the 3
and 9 o’clock position of the pipe) to prevent signal attenuation caused by sediment on
the bottom of the pipe or air bubbles and air pockets in the top of the pipe.