![SUPMEA SUP-1158-J Скачать руководство пользователя страница 15](http://html1.mh-extra.com/html/supmea/sup-1158-j/sup-1158-j_user-manual_1406088015.webp)
Chapter 1 Installation
www.supmea.com
- 6 -
Backspace or delete characters to the left.
And
Return to the last menu or to open the
next menu. Acts as "+" and "-" functions when entering numbers.
Select a menu. Press this key first, input two menu numbers and then enter
the selected menu. For example, to input a pipe outside diameter, press
keys, where "11" is the window ID to display the parameter for
pipe outside diameter.
1.5. Key operation
The flow meter adopts the window software design to consolidate or subdivide all
of the parameters entered, the instrument setup and measurement result displays
into more than 100 independent windows. The operator can input parameters,
modify settings or display measurement results by "visiting" a specific window.
These windows are arranged by 2-digit serial numbers (including "+" sign) from
00~99, then to +0, +1, etc. Each window serial number, or so-called window ID
code, has a defined meaning. For example, Window M11 indicates the parameter
input for pipe outside diameter, while Window M25 indicates the mounting spacing
between the transducers, etc. (Refer – Windows Display Explanations).
The keypad shortcut to visit a specific window is to press the
key at any time,
then input the 2-digit window ID code. For example, to input or check the pipe