16
5.0 Operation
5.1.2 Setting the Test Pressure
If the
40002 Servo_Mode
register has been set to
T
est Pressure
mode
. The operator should verify and if
necessary set the test pressure prior to testing.
Test Pressure Register
Register
Function
Data Type
Read/
Write
Value
Units
Scale
40035-36
TestPressureSetFloat
Float-32
R/W
Floating point value for test pressure setpoint InH
2
0
None
Register
40035-36 TestPressureSetFloat
provides the status (state) that the flowbench test pressure
is currently in when read. It also allows for the ability to change (set) states when this register receives
commands. The register can be set to the desired test pressure in the
unit of measurement
(InH
2
O).
Example:
25
= A test pressure of 25" H
2
0
5.1.3 Setting the Flow Rate
If the
40002 Servo_Mode
register has been set to
Flow Rate
mode
. The operator should verify and set the
flow rate prior to testing.
Flow Rate Register
Register
Function
Data Type
Read/
Write
Value
Units
Scale
40037-38
FlowRateSetFloat
Float-32
R/W
Floating point value for flow rate setpoint
CFM
None
Register
40037-38 FlowRateSetFloat
provides the status (state) that the flowbench flow rate is currently in
when read. It also allows for the ability to change (set) states when this register receives commands. The
register can be set to the desired flow rate in the
unit of measurement
(CFM).
Example:
100
= A flow rate of 100 cfm