14
It has the function of calculating intraocular lens and the name of calculating formula
should be given in software settings and attached files.
(2) Appearance and structural requirements
The shell of the diagnostic instrument should be free from mechanical damage and
rust. The words and marks on the panel should be clearly visible and firm.
The plastic parts of the diagnostic instrument should be free from foaming, cracking,
deformation and overflow of fillings.
The operation and adjusting mechanism of the diagnostic instrument should be
flexible, reliable and the fastener should not be loosened.
The front side of the B-scan probe of the diagnostic instrument should be marked
correctly to indicate the scanning plane, and the marking should be on the scanning
plane.
(3) Foot pedal
The pedal switch shall meet the requirements of MEDICAL FOOT SWITCH.
(4) Safety requirements
General safety shall meet the requirements of EN 60601-1:2006/ A1:2013 or IEC
60601-1:2005 /A1:2012 and special safety shall meet the requirements of IEC/EN
60601-2-37 2015.
(5) EMC requirements
EMC shall comply with the relevant provisions of EN 60601-1-2:2015/ IEC
60601-1-2:2014 and IEC/EN 60601-2-37 2015.
(6) Environmental test requirements
The measuring instrument should meet the requirements of Climate Environment
Group II and Mechanical Environment Group II in
EN ISO
14710-2009
2.3 Theory
A-scan theory
Содержание SK-3000A
Страница 12: ...11 1 12 Label This label information is stick on main body of Ophthalmic ultrasound ...
Страница 28: ...27 4 5 P scan Test Interface 1 Eye Switch left eye and right eye ...
Страница 30: ...29 4 6 B scan Test Interface 1 Eye Switch left eye and right eye ...
Страница 35: ...34 ...
Страница 41: ...40 2 Click Properties 3 Double click Internet protocol version 4 ICP IPv4 ...