![Suez Sievers UPW Boron Analyzer Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/suez/sievers-upw-boron-analyzer/sievers-upw-boron-analyzer_operation-and-maintenance-manual_1389722025.webp)
SUEZ © 2018
DLM 71000-03 EN
Rev. B
Chapter 2. System Description
System Specifications
operation
Range
Limit of Quantification (LOQ)
Limit of Detection (LOD)
Accuracy
Precision
Minimum Sample Resistivity
Sample Temperature
Sample Flow Rate
Ambient Temperature
Sample Pressure
Calibration Stability
Sample Particulates
Waste
Altitude
Inputs
Outputs
Transient Overvoltage
Dimensions
Weight
Overall Site Dimensions
Minimum Clearances (for temperature,
humidity control, and access)
consumables
Boron Reagent
Reagent Column
Zero Column
Peristaltic Pump Tubing
0.015 to 20 ppb B
0.05 ppb B
0.015 ppb B
± 0.01 ppb B < 0.05 ppb B; ±15%
0.05 ppb B
± 3% or 0.005 ppb B, whichever is greater
15 M
-cm
15–40 °C
0.1–0.5 L/min
10–40 °C
20–100 psig (138–689 kPa)
Typically 12 months
15-micron pre-filtration
Gravity drain (no back pressure)
Designed for use at altitudes up to 2000 m
5 isolated binary inputs
4 isolated 4-20 mA outputs; 1 RS-232 (isolated); 1 universal
parallel printer port; 2 alarm outputs; 4 isolated binary outputs
Installation (Overvoltage) Category II (Instrument Environment Pollution
Degree 2 assumed for purposes of testing and judging insulation integrity)
53 H x 44 W x 31 D cm (20 5/8 x 17 1/4 x 12 1/4 in)
21.75 kg (48 lb.)
Height: 97 cm; width: 93 cm; depth: 62 cm
Front: 38 cm; back 1 cm; sides: 30 cm each side; top: 30 cm;
bottom: 30 cm from Analyzer (or shelf on which it sits) to the gravity drain
90 days at 12-minute sampling rate; 45 days at 6.5-minute sampling rate
12 months
12 months
12 months
instrument
Содержание Sievers UPW Boron Analyzer
Страница 4: ...SUEZ 2018 4 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 16: ...SUEZ 2018 16 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 24: ...SUEZ 2018 24 of 104 DLM 71000 03 EN Rev B Chapter 1 Introduction THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 64: ...SUEZ 2018 64 of 104 DLM 71000 03 EN Rev B Chapter 5 System Configuration THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 96: ...SUEZ 2018 96 of 104 DLM 71000 03 EN Rev B Chapter 8 Maintenance THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 100: ...SUEZ 2018 100 of 104 DLM 71000 03 EN Rev B Chapter 9 Message and Error Codes THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 103: ...Notes SUEZ 2018 103 of 104 DLM 71000 03 EN Rev B 104...
Страница 104: ...Notes SUEZ 2018 104 of 104 DLM 71000 03 EN Rev B 104...