To replace the fan filter
1. Use a Phillips head screwdriver to remove the two screws in the cover plate at
the bottom of the Analyzer front. See
. Remove the
cover plate.
2.
Reach in and extract the filter.
3.
Re-install a new filter by sliding it back into place.
4.
Re-install the cover and tighten the two screws.
Entering New Consumables Installation Dates
When replacing a consumable, enter the date of installation on the
Consumables
screen to
enable the Analyzer to keep an accurate record of usage and issue reminders.
1. On the
Settings
screen, select the
Consumables
tab.
2.
In the applicable area (
UV Lamp
,
Pump Tubing
,
Resin Cartridge
), click one of
the following buttons to use the calendar for entering the date the consumable
was installed. The installation date, applicable expiration date, and number of
days remaining for the life of the consumables automatically appears to the left
of the button.
•
Install UV Lamp
•
Install Pump Tubing
•
Install Resin Cartridge
3.