Warnings and Error Messages
This section details some of the warnings and error messages that have troubleshooting
steps that can be performed by the user. See Appendix C of this manual for a full list of
warnings and error messages.
3100—U
NEXPECTED
DI L
OOP
LEVEL
This warning can occur if the water level in the DI Reservoir is low. To troubleshoot:
1. Check the water level in the DI reservoir. The water level should touch the FULL
mark on the outside of the reservoir:
2. If the water level is low, add water by following the instructions in
.
3. If the DI reservoir is full when this error occurs, power cycle the Analyzer. Stop
any analysis and click the Power Off
P
button in the Analyzer’s firmware, then
turn
off
the power switch. Wait 30 seconds, then turn on the power switch.
4. If the error persists, contact your local service provider or Technical Support for
further assistance.
3102—DI
FLOW
ON
IC/TC
CHANNEL
NOT
DETECTED
This warning can occur if the flow in the IC/TC channel is low. To troubleshoot, check
the flow through the restrictor tube by following the steps in
.