Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
235
SUEZ © 2022
TOC Multi-Point Calibration Worksheet
Company Name
Analyzer Serial Number
Firmware Version
Analyzer Location
Standard Lot # and Expiration
P
ASS
/F
AIL
C
RITERIA
:
Cal Blank 1 IC RSD ≤ 3.00%
Cal Blank 1 TC RSD ≤ 3.00%
1.5 ppm TOC RSD ≤ 3.00%
1 ppm TOC RSD ≤ 3.00%
500 ppb TOC RSD ≤ 3.00%
Change in IC Slope ± 20.0%
Change in TC Slope ± 20.0%
R-squared ≥ 0.988
The attached data will indicate a
Passed
or
Failed
condition.
Results:
Passed
Failed
Performed by:
Date:
Reviewed by:
Date:
Verified by: