![Suez Sievers M500e Скачать руководство пользователя страница 234](http://html1.mh-extra.com/html/suez/sievers-m500e/sievers-m500e_operation-and-maintenance-manual_1389721234.webp)
L
OW
L
EVEL
TOC S
INGLE
P
OINT
C
ALIBRATION
W
ORKSHEET
.............................
Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
234
SUEZ © 2022
Low Level TOC Single Point Calibration Worksheet
Company Name
Analyzer Serial Number
Firmware Version
Analyzer Location
Standard Lot # and Expiration
P
ASS
/F
AIL
C
RITERIA
:
Cal Blank 1 IC RSD ≤ 3.00%
Cal Blank 1 TC RSD ≤ 3.0%
250 ppb TOC RSD ≤ 3.0%
Change in IC Slope ± 20.0%
Change in TC Slope ± 20.0%
The attached data will indicate a
Passed
or
Failed
condition.
Results:
Passed
Failed
Performed by:
Date:
Reviewed by:
Date:
Verified by:
Date: