C
ALIBRATION
P
REPARATIONS
.............................
Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
192
SUEZ © 2022
C
ALIBRATION
P
REPARATIONS
Saving System Settings
Prior to performing any calibration procedure or system protocol, save the current
system settings, in the event they need to be re-loaded. (Consider also printing the
settings for future reference.)
To save the system settings
1. On the
Instrument Settings
s
screen >
Instrument Settings
tab, select
Analyzer Settings
.
2.
Click
Export Settings
.
3.
The next screen shows file type, file name, and path options for the system
settings export file.
4.
Select the Export File Type from the list of options on the right side of the
screen.
5.
The export file path chosen during system setup is automatically selected.
NOTE:
To change this path, go to the
Settings
s
screen >
System
Setup
tab, and select
Export Preferences
.