The TC/IC Conductivity Autozero protocol determines the IC and TC cell offsets. The
Analyzer can be set to automatically perform a TC/IC Conductivity Autozero at a specific
interval, or the user can run the protocol manually.
1. If the Analyzer is taking measurements, click
Stop Analysis
.
2. On the
System Protocols
p
screen, select the
Calibration
tab.
3. Select
TC/IC Conductivity Autozero
, then click
Setup
.
4. The
Setup TC/IC Conductivity Autozero
dialog box appears.
5. If it is not already enabled, select the
Enable Scheduled Autozero
check box.