SF-8200 Operation Manual
52
/
76
2.3 CREATE A CALIBRATION CURVE
1) Click the "New" button to display the setting interface for adding calibration test items on the right;
2) Set the calibration test
In the project drop-down box, select the test items to be calibrated;
Select the calibration method, there are two ways of "automatic dilution and analysis" and
"manual input"
3) Automatic dilution and analysis
New calibration curve, select "automatic dilution and analysis" mode. In the pop-up "Select Cali-
brator batch number", select the batch number of the calibrator loaded;
Click "OK" button to generate calibration task, and automatically jump to "test" interface, click on
the function panel area "start" button to start the test task;
①
②