Installation and connection

5.3

Install device

Process pipework has been properly prepared.
Process pipework has been secured against unintentional
opening with shut-off values.
Installation position vertical, direction of

fl

ow from bottom

to top.

WARNING

Risk of injury and poisoning due to medium spraying out.

Ensure that in case of a break in the measuring tube, no
hazards result from escaping media.
Do not use measuring tube made of PVC for gaseous
media.

WARNING

Risk of injury and poisoning due to medium spraying out.

Use personal protective equipment when carrying out any
work on the

fi

tting.

1. Ensure that measuring tube (6) and

fl

oat (4) are clean and

free of foreign matter.

2. Remove transport lock (holds the

fl

oat).

3. Ensure that the

fl

oat (4) can move freely.

5.3.2

Connection with union nut and insert

1. Unscrew both union nuts (1)

(

Figure Layout, Page 5 ).

Secure

fl

oat (heavy) from falling

2. Push union nuts (1) onto spigot of process pipes.

Check mounting direction.

3. Weld union end (2) of device to spigot of process pipes.
4. Check

fi

t of O-rings.

5. Connect device to the process pipework. Tighten union

nuts (1) by hand only.

5.3.3

Connection with

fl

ange

1. Prepare pipe ends according to connection type.
2. Radially push the

fi

tting and

fl

at seal between the

fl

ange

ends.

3. Bolt

fi

tting and

fl

ange with

fl

ange screws, nuts and wa-

shers.
While

doing

so,

observe

tightening

torques:

(

9.2 Tightening torques, Page 12).

5.3.4

If necessary: Attach special scale

1

2

1. With a pen, mark the lowest reading point of the existing

scale on the measuring tube (1).

2. Pull current scale off.
3. Stick new scale on so that the marking on the measuring

tube matches with the lowest reading point (2).

5.4

Performing the hydrostatic test

Perform hydrostatic test using neutral medium, e.g. water.

1. Pressurize the device, ensuring

Test pressure < 1.5 x P

N

(Nominal pressure)

Test pressure < P

N

+ 5 bar

Test pressure < permissible system pressure

2. Check that the device is not leaking.

8

DFM 165 – 350

BA-2016.08.09 EN

300 458

Содержание DFM 350

Страница 1: ... No 300 458 TR MA DE Rev002 ASV Stübbe GmbH Co KG Hollwieser Straße 5 32602 Vlotho Germany Phone 49 0 5733 799 0 Fax 49 0 5733 799 5000 E mail contact asv stuebbe de Internet www asv stuebbe com We reserve the right to make technical changes Read carefully before use Save for future use ...

Страница 2: ...ntrol devices recommended 7 5 3 Install device 8 5 3 2 Connection with union nut and insert 8 5 3 3 Connection with flange 8 5 3 4 If necessary Attach special scale 8 5 4 Performing the hydrostatic test 8 6 Operation 9 6 1 Commissioning 9 6 2 Read measurement 9 7 Maintenance 10 7 1 Servicing 10 7 2 Maintenance 10 7 2 1 Removing fitting 10 7 3 Replacement parts and return 10 8 Troubleshooting 11 8 ...

Страница 3: ...is manual and the other applicable documents especially all safety instructions and warnings 1 2 Other applicable documents Resistance lists Resistance of materials used to chemicals www asv stuebbe de pdf_resistance 300051 pdf Data sheet Technical specifications conditions of operation www asv stuebbe de pdf_datasheets 300464 pdf CE declaration of conformity Conformity with standards www asv stue...

Страница 4: ...where the pump is operated Make personal protective equipment available Qualified personnel Make sure all personnel tasked with work on the device have read and understood this manual and all other appli cable documents especially the safety maintenance and repair information before they start any work Organize responsibilities areas of competence and the supervision of personnel The following wor...

Страница 5: ...askets 3 2 Description The device measures the flow of a liquid or gaseous medium The medium flows through the flowmeter vertically from bot tom to top The float is lifted by the flow forces On the upper reading edge largest diameter the flow quantity in volume or mass can be read on the measuring tube scale The displayed value only applies for the medium and medium temperature for which the devic...

Страница 6: ...storage Store the device properly 1 Make sure the storage room meets the following condi tions Dry Frost free Vibration free Not in direct sunlight Storage temperature 10 C to 60 C 2 Device should preferably be stored in the original packag ing 4 4 Disposal Plastic parts can be contaminated by poisonous or radioac tive media to such an extent that cleaning will not be suffi cient WARNING Risk of p...

Страница 7: ...om mended to avoid vibration of the float DN DN device Length 5 7 x DN device Length 10 x DN device for large differences between nominal diameters between pipe and device 4 If necessary For gaseous media equip with vent valve at highest point of inlet and outlet pipe For condensing gaseous equip with vent valve at low est point of inlet and outlet pipe 5 Allocate control valves after device 6 All...

Страница 8: ...ng 2 Push union nuts 1 onto spigot of process pipes Check mounting direction 3 Weld union end 2 of device to spigot of process pipes 4 Check fit of O rings 5 Connect device to the process pipework Tighten union nuts 1 by hand only 5 3 3 Connection with flange 1 Prepare pipe ends according to connection type 2 Radially push the fitting and flat seal between the flange ends 3 Bolt fitting and flange...

Страница 9: ...s in inlet and outlet line 2 Ensure that the float can float freely and measures a plau sible value 6 2 Read measurement For the measurement range 2 500 25 000 the reading edge is below the float fixture largest diameter 1 1 Reading edge Read measurement using face 1 of float 300 458 BA 2016 08 09 EN DFM 165 350 9 ...

Страница 10: ...eposits or changes to surface Tab 3 Servicing activities Perform maintenance tasks according to the table 7 2 Maintenance WARNING Risk of injury during disassembly Wear protective gloves components can be very sharp edged due to wear or damage 7 2 1 Removing fitting 1 Ensure that System is empty System has been flushed System is depressurized System has cooled down System is secured against being ...

Страница 11: ...ht defective Check weight of float Ask for weight from manufacturer If weight differs remove device insert new float and re install device 8 1 Clean measuring tube and float Page 11 No display only for measurement sensor ZE3000 ZE3075 or limit sensor Float installed incorrectly Ensure that a float with magnet is used Tab 4 Troubleshooting 8 1 Clean measuring tube and float System is empty System h...

Страница 12: ...tor PE Union nut PVC PP PVDF Tab 5 Mechanical specifications 9 1 2 Dimensions Dimensions Data sheet 9 2 Tightening torques Tightening torque Nm for the sizes d mm Description 16 20 25 32 40 50 63 75 Flange GFK 5 7 10 15 20 25 32 37 Flange PP steel 10 15 15 20 25 35 40 Tab 6 Tightening torques 1 Greased housing screws hex screws hex nuts 9 3 Accessories Description Flow data sensor ZE 3000 magnet a...

Отзывы: