Preventive maintenance
Remove product from service before you perform the preventive maintenance inspection. Check all items listed during
annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance
checks more often based on your level of product usage. Service only by qualified personnel.
WARNING
Risk of electric shock. Disconnect power before servicing or cleaning the
T/Pump
.
Note:
Change the water monthly or more often depending on use.
Inspect the following items:
Line cord, the plug, and the housing for cuts or cracked insulation and quality of the strain relief, if damaged
replace
Membrane panel for cuts or cracks, if damaged replace
Plastic parts on the pump for cracks
Connectors for cracks and missing O-rings, replace connectors if necessary
Hose connections and connectors function properly
Pads and hoses are free from tears, cracks, and water leaks
Refer to the
T/Pump
Maintenance Manual Functional Checks section that and complete during preventive maintenance.
Inspection
Value
Okay
Action Needed
Action Taken
Measure current
leakage <100 µA
Product serial number:
Completed by:
Date:
Содержание T/Pump Series
Страница 2: ...sample text ...
Страница 4: ...sample text ...
Страница 12: ...English EN 1 8 101515 REV E www stryker com Introduction Serial number location ...
Страница 28: ...sample text ...
Страница 30: ...sample text ...
Страница 38: ...Français FR 2 8 101515 REV E www stryker com Introduction Emplacement du numéro de série ...
Страница 53: ......
Страница 54: ...2017 10 101515 REV E www stryker com Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA ...